Intracoronary Optical Coherence Tomography 2018: Current Status and Future Directions Fate of post-procedural malapposition of everolimus-eluting polymeric bioresorbable scaffold and everolimus-eluting cobalt chromium metallic stent in human coronary arteries: sequential assessment with optical coherence tomography in ABSORB Japan trial Optical coherence tomography findings: insights from the “randomised multicentre trial investigating angiographic outcomes of hybrid sirolimus-eluting stents with biodegradable polymer compared with everolimus-eluting stents with durable polymer in chronic total occlusions” (PRISON IV) trial Combining IVUS and Optical Coherence Tomography for More Accurate Coronary Cap Thickness Quantification and Stress/Strain Calculations: A Patient-Specific Three-Dimensional Fluid-Structure Interaction Modeling Approach Comparison of Stent Expansion Guided by Optical Coherence Tomography Versus Intravascular Ultrasound: The ILUMIEN II Study (Observational Study of Optical Coherence Tomography [OCT] in Patients Undergoing Fractional Flow Reserve [FFR] and Percutaneous Coronary Intervention) Optical coherence tomography imaging during percutaneous coronary intervention impacts physician decision-making: ILUMIEN I study Intracoronary Optical Coherence Tomography-Derived Virtual Fractional Flow Reserve for the Assessment of Coronary Artery Disease Optical coherence tomography compared with intravascular ultrasound and with angiography to guide coronary stent implantation (ILUMIEN III: OPTIMIZE PCI): a randomised controlled trial Randomized comparison of stent strut coverage following angiography- or optical coherence tomography-guided percutaneous coronary intervention Vascular response and healing profile of everolimus-eluting bioresorbable vascular scaffolds for percutaneous treatment of chronic total coronary occlusions: A one-year optical coherence tomography analysis from the GHOST-CTO registry

Clinical TrialPublished on 4 August 2017

JOURNAL:EuroIntervention. Article Link

Optical coherence tomography findings: insights from the “randomised multicentre trial investigating angiographic outcomes of hybrid sirolimus-eluting stents with biodegradable polymer compared with everolimus-eluting stents with durable polymer in chronic total occlusions” (PRISON IV) trial

Teeuwen K, Spoormans EM, Bennett J et al. Keywords: drug-eluting stent; optical coherence tomography; stable angina; Chronic coronary total occlusion

ABSTRACT

AIMS - The PRISON IV trial investigated the next-generation sirolimus-eluting stent (SES) with ultra-thin struts and biodegradable polymer against the second-generation everolimus-eluting stent (EES) with thin struts and durable polymer in patients with successfully recanalised chronic total occlusions (CTO). In this study, we examined the secondary optical coherence tomography endpoints.


METHODS AND RESULTS - The main PRISON IV trial randomised 330 patients to either SES or EES. At nine months, 281 (85%) patients underwent repeat angiography. Of these, 60 consecutive patients received optical coherence tomography divided over both stent groups. The mean number of struts analysed was 750±337 and 633±358 in SES and EES patients, respectively (p=0.07). The minimal lumen area, minimal stent area, maximal neointima area and neointimal thickness were comparable between the groups (4.8±2.1 and 4.4±1.5 mm2; 5.3±1.8 and 5.3±1.4 mm2; 2.5±2.0 and 2.2±1.5 mm2; 0.7±1.7 and 0.4±0.2 mm). The percentage of uncovered struts was higher with EES (6.2±7.5% and 11.9±13.4%, p=0.04), whereas the percentage of malapposed struts and mean number of coronary evaginations were significantly higher with SES (2.9±4.0% and 1.2±2.4%, p=0.02; 18.5±17.7 and 5.3±3.1, p=0.004).


CONCLUSIONS - The optical coherence tomography findings of this substudy demonstrated improved strut coverage with ultra-thin strut SES with bioresorbable polymer compared to thin-strut EES with durable polymer in CTO. On the other hand, SES showed a higher rate of stent strut malappositon and coronary evaginations. The clinical relevance of these findings remains to be demonstrated.