CBS 2019
CBSMD教育中心
English

血流储备分数

科研文章

荐读文献

Fractional Flow Reserve-Guided Multivessel Angioplasty in Myocardial Infarction Impact of Percutaneous Revascularization on Exercise Hemodynamics in Patients With Stable Coronary Disease Prognostic Implication of Thermodilution Coronary Flow Reserve in Patients Undergoing Fractional Flow Reserve Measurement Meta-Analysis of Death and Myocardial Infarction in the DEFINE-FLAIR and iFR-SWEDEHEART Trials Fractional Flow Reserve–Guided PCI for Stable Coronary Artery Disease Retrospective Comparison of Long-Term Clinical Outcomes Between Percutaneous Coronary Intervention and Medical Therapy in Stable Coronary Artery Disease With Gray Zone Fractional Flow Reserve - COMFORTABLE Retrospective Study High-Resolution Cardiac Magnetic Resonance Imaging Techniques for the Identification of Coronary Microvascular Dysfunction Diagnostic accuracy of fractional flow reserve from anatomic CT angiography Lesion-Specific and Vessel-Related Determinants of Fractional Flow Reserve Beyond Coronary Artery Stenosis Diagnosis of ischemia-causing coronary stenoses by noninvasive fractional flow reserve computed from coronary computed tomographic angiograms. Results from the prospective multicenter DISCOVER-FLOW

Original Research2019 Jan 22;139(4):477-484.

JOURNAL:Circulation. Article Link

Accuracy of Fractional Flow Reserve Derived From Coronary Angiography

Fearon WF, Achenbach S, FAST-FFR Study Investigators et al. Keywords: coronary artery disease; coronary circulation; fractional flow reserve, myocardial

FULL TEXT PDF

BACKGROUND - Measuring fractional flow reserve (FFR) with a pressure wire remains underutilized because of the invasiveness of guide wire placement or the need for a hyperemic stimulus. FFR derived from routine coronary angiography (FFRangio) eliminates both of these requirements and displays FFR values of the entire coronary tree. The FFRangio Accuracy versus Standard FFR (FAST-FFR) study is a prospective, multicenter, international trial with the primary goal of determining the accuracy of FFRangio.

 

METHODS - Coronary angiography was performed in a routine fashion in patients with suspected coronary artery disease. FFR was measured in vessels with coronary lesions of varying severity using a coronary pressure wire and hyperemic stimulus. Based on angiograms of the respective arteries acquired in 2 different projections, on-site operators blinded to FFR then calculated FFRangio using proprietary software. Coprimary end points were the sensitivity and specificity of the dichotomously scored FFRangio for predicting pressure wire-derived FFR using a cutoff value of 0.80. The study was powered to meet prespecified performance goals for sensitivity and specificity.

 

RESULTS - Ten centers in the United States, Europe, and Israel enrolled a total of 301 subjects and 319 vessels meeting inclusion/exclusion criteria which were included in the final analysis. The mean FFR was 0.81 and 43% of vessels had an FFR0.80. The per-vessel sensitivity and specificity were 94% (95% CI, 88% to 97%) and 91% (86% to 95%), respectively, both of which exceeded the prespecified performance goals. The diagnostic accuracy of FFRangio was 92% overall and remained high when only considering FFR values between 0.75 to 0.85 (87%). FFRangio values correlated well with FFR measurements ( r=0.80, P<0.001) and the Bland-Altman 95% confidence limits were between -0.14 and 0.12. The device success rate for FFRangio was 99%.

 

CONCLUSIONS - FFRangio measured from the coronary angiogram alone has a high sensitivity, specificity, and accuracy compared with pressure wire-derived FFR. FFRangio has the promise to substantially increase physiological coronary lesion assessment in the catheterization laboratory, thereby potentially leading to improved patient outcomes.

 