Technical aspects of the culotte technique Angiography Alone Versus Angiography Plus Optical Coherence Tomography to Guide Percutaneous Coronary Intervention Outcomes From the Pan-London PCI Cohort Anatomical and Functional Computed Tomography for Diagnosing Hemodynamically Significant Coronary Artery Disease: A Meta-Analysis Characteristics of abnormal post-stent optical coherence tomography findings in hemodialysis patients The Relation Between Optical Coherence Tomography-Detected Layered Pattern and Acute Side Branch Occlusion After Provisional Stenting of Coronary Bifurcation Lesions Spontaneous Coronary Artery Dissection: Pathophysiological Insights From Optical Coherence Tomography Optical coherence tomography versus intravascular ultrasound to evaluate coronary artery disease and percutaneous coronary intervention Incidence of Adverse Events at 3 Months Versus at 12 Months After Dual Antiplatelet Therapy Cessation in Patients Treated With Thin Stents With Unprotected Left Main or Coronary Bifurcations Double-Kiss-Crush Bifurcation Stenting: Step-by-Step Troubleshooting Clinical use of intracoronary imaging. Part 1: guidance and optimization of coronary interventions. An expert consensus document of the European Association of Percutaneous Cardiovascular Interventions: Endorsed by the Chinese Society of Cardiology

Original ResearchVolume 13, Issue 6, March 2020

JOURNAL:JACC Cardiovasc Interv. Article Link

Feasibility of Coronary Access and Aortic Valve Reintervention in Low-Risk TAVR Patients

T Rogers, BC Greenspun, G Weissman et al. Keywords: coronary access; coronary artery obstruction; PCI; TAVR; valve-in-valve

ABSTRACT


OBJECTIVES - The aim of this study was to evaluate the feasibility of coronary access and aortic valve reintervention in low-risk patients undergoing transcatheter aortic valve replacement (TAVR) with a balloon-expandable transcatheter heart valve (THV).

 

BACKGROUND - Younger, low-risk TAVR patients are more likely than older, higher risk patients to require coronary angiography, percutaneous coronary intervention, or aortic valve reintervention, but their THVs may impede coronary access and cause coronary obstruction during TAVR-in-TAVR.

 

METHODS - The LRT (Low Risk TAVR) trial (NCT02628899) enrolled 200 subjects with symptomatic severe aortic stenosis to undergo TAVR using commercially available THVs. Subjects who received balloon-expandable THVs and who had 30-day cardiac computed tomographic scans were included in this study. In a subgroup, the feasibility of intentional THV crimping on the delivery catheter to pre-determine commissural alignment was tested.

 

RESULTS - In the LRT trial, 168 subjects received balloon-expandable THVs and had 30-day cardiac computed tomographic scans, of which 137 were of adequate image quality for analysis. The most challenging anatomy for coronary access (THV frame above and commissural suture post in front of a coronary ostium) was observed in 9% to 13% of subjects. Intentional THV crimping did not appear to meaningfully affect commissural alignment. The THV frame extended above the sinotubular junction in 21% of subjects, and in 13%, the distance between the THV and the sinotubular junction was <2 mm, signifying that TAVR-in-TAVR may not be feasible without causing coronary obstruction.

 

CONCLUSIONS - TAVR may present challenges to future coronary access and aortic valve reintervention in a substantial number of low-risk patients.