Optical Coherence Tomography

Abstract

Recommended Article

Optical Coherence Tomography Findings in Patients With Coronary Stent Thrombosis: A Report of the PRESTIGE Consortium (Prevention of Late Stent Thrombosis by an Interdisciplinary Global European Effort) Coronary Optical Coherence Tomography and Cardiac Magnetic Resonance Imaging to Determine Underlying Causes of Myocardial Infarction With Nonobstructive Coronary Arteries in Women A new optical coherence tomography-based calcium scoring system to predict stent underexpansion Superficial Calcium Fracture After PCI as Assessed by OCT Drug-eluting stent implantation in patients with acute coronary syndrome - the Activity of Platelets after Inhibition and Cardiovascular Events: Optical Coherence Tomography (APICE OCT) study Lipid-rich plaque and myocardial perfusion after successful stenting in patients with non-ST-segment elevation acute coronary syndrome: an optical coherence tomography study Nonculprit Lesion Plaque Morphology in Patients With ST-Segment–Elevation Myocardial Infarction: Results From the COMPLETE Trial Optical Coherence Tomography Substudys Characteristics of stent thrombosis in bifurcation lesions analysed by optical coherence tomography

Original Research2016 Jan 1;87(1):E9-E14.

JOURNAL:Catheter Cardiovasc Interv. Article Link

Optimal threshold of postintervention minimum stent area to predict in-stent restenosis in small coronary arteries: An optical coherence tomography analysis

Matsuo Y, Kubo T, Aoki H et al. Keywords: everolimus-eluting stent; minimum stent area; optical coherence tomography; small vessel

ABSTRACT


OBJECTIVES - The aim of this study was to determine the best threshold of postintervention minimum stent area (MSA) assessed by optical coherence tomography (OCT) to predict long-term in-stent restenosis (ISR) for 2.5 mm-diameter everolimus-eluting stents (EES).

 

BACKGROUND - Percutaneous coronary intervention (PCI) for small coronary arteries remains challenging. Stent underexpansion is a strong predictor of late ISR.

 

METHODS - We performed a retrospective analysis of 69 lesions in 69 patients undergoing PCI with 2.5 mm-diameter stents using OCT for the assessment of postintervention MSA and subsequent 9-month angiographic follow-up.

 

RESULTS - The rates of angiographic ISR and target lesion revascularization were 7.2% and 1.4%. The postintervention OCT-MSA of EES < 3.5 mm(2) for predicting ISR yielded a sensitivity of 80%, specificity of 71%, positive predictive value of 18%, and negative predictive value of 98%. There was a marginally significant trend between increasing MSA quartiles and decreasing ISR rate (P for trend = 0.07).

 

CONCLUSIONS - Postintervention OCT-MSA of 3.5 mm(2) best predicted 9-month ISR following PCI with 2.5-mm-diameter EES. Further large, prospective, observational studies are warranted that validate this result.