Abstract

Recommended Article

Cutoff Value and Long-Term Prediction of Clinical Events by FFR Measured Immediately After Implantation of a Drug-Eluting Stent in Patients With Coronary Artery Disease: 1- to 3-Year Results From the DKCRUSH VII Registry Study Prognostic Value of Fractional Flow Reserve Measured Immediately After Drug-Eluting Stent Implantation Clinical value of post-percutaneous coronary intervention fractional flow reserve value: A systematic review and meta-analysis Prognostic Implication of Thermodilution Coronary Flow Reserve in Patients Undergoing Fractional Flow Reserve Measurement Anatomical plaque and vessel characteristics are associated with hemodynamic indices including fractional flow reserve and coronary flow reserve: A prospective exploratory intravascular ultrasound analysis Clinical Significance of Concordance or Discordance Between Fractional Flow Reserve and Coronary Flow Reserve for Coronary Physiological Indices, Microvascular Resistance, and Prognosis After Elective Percutaneous Coronary Intervention FFR-guided multivessel stenting reduces urgent revascularization compared with infarct-related artery only stenting in ST-elevation myocardial infarction: A meta-analysis of randomized controlled trials Correlation between frequency-domain optical coherence tomography and fractional flow reserve in angiographically-intermediate coronary lesions

Original Research2019 Jan 22;139(4):477-484.

JOURNAL:Circulation. Article Link

Accuracy of Fractional Flow Reserve Derived From Coronary Angiography

Fearon WF, Achenbach S, FAST-FFR Study Investigators et al. Keywords: coronary artery disease; coronary circulation; fractional flow reserve, myocardial

FULL TEXT PDF

BACKGROUND - Measuring fractional flow reserve (FFR) with a pressure wire remains underutilized because of the invasiveness of guide wire placement or the need for a hyperemic stimulus. FFR derived from routine coronary angiography (FFRangio) eliminates both of these requirements and displays FFR values of the entire coronary tree. The FFRangio Accuracy versus Standard FFR (FAST-FFR) study is a prospective, multicenter, international trial with the primary goal of determining the accuracy of FFRangio.

 

METHODS - Coronary angiography was performed in a routine fashion in patients with suspected coronary artery disease. FFR was measured in vessels with coronary lesions of varying severity using a coronary pressure wire and hyperemic stimulus. Based on angiograms of the respective arteries acquired in 2 different projections, on-site operators blinded to FFR then calculated FFRangio using proprietary software. Coprimary end points were the sensitivity and specificity of the dichotomously scored FFRangio for predicting pressure wire-derived FFR using a cutoff value of 0.80. The study was powered to meet prespecified performance goals for sensitivity and specificity.

 

RESULTS - Ten centers in the United States, Europe, and Israel enrolled a total of 301 subjects and 319 vessels meeting inclusion/exclusion criteria which were included in the final analysis. The mean FFR was 0.81 and 43% of vessels had an FFR0.80. The per-vessel sensitivity and specificity were 94% (95% CI, 88% to 97%) and 91% (86% to 95%), respectively, both of which exceeded the prespecified performance goals. The diagnostic accuracy of FFRangio was 92% overall and remained high when only considering FFR values between 0.75 to 0.85 (87%). FFRangio values correlated well with FFR measurements ( r=0.80, P<0.001) and the Bland-Altman 95% confidence limits were between -0.14 and 0.12. The device success rate for FFRangio was 99%.

 

CONCLUSIONS - FFRangio measured from the coronary angiogram alone has a high sensitivity, specificity, and accuracy compared with pressure wire-derived FFR. FFRangio has the promise to substantially increase physiological coronary lesion assessment in the catheterization laboratory, thereby potentially leading to improved patient outcomes.

 