CBS 2019
CBSMD教育中心
中 文

分叉支架

Abstract

Recommended Article

Anatomical Attributes of Clinically Relevant Diagonal Branches in Patients with Left Anterior Descending Coronary Artery Bifurcation Lesions Feasibility and efficacy of the ultrashort side branch dedicated balloon in coronary bifurcation stenting Three-Year Outcomes of the DKCRUSH-V Trial Comparing DK Crush With Provisional Stenting for Left Main Bifurcation Lesions Metabolic Interactions and Differences between Coronary Heart Disease and Diabetes Mellitus: A Pilot Study on Biomarker Determination and Pathogenesis Tips of the dual-lumen microcatheter-facilitated reverse wire technique in percutaneous coronary interventions for markedly angulated bifurcated lesions Double-Kissing Culotte Technique for Coronary Bifurcation Stenting - Technical evaluation and comparison with conventional double stenting techniques A Randomized Trial Evaluating Online 3-Dimensional Optical Frequency Domain Imaging-Guided Percutaneous Coronary Intervention in Bifurcation Lesions Randomized trial of simple versus complex drug-eluting stenting for bifurcation lesions: the British Bifurcation Coronary Study: old, new, and evolving strategies

Original Research2020 Jul 14;EIJ-D-20-00334.

JOURNAL:Eurointervention. Article Link

Feasibility and efficacy of the ultrashort side branch dedicated balloon in coronary bifurcation stenting

Y Murasato, M Nishihara, T Mori et al. Keywords: kissing balloon inflation vs. glider balloon; stent configuration; high-angled bifurcation; coronary bifurcation stenting

ABSTRACT

AIMS - We sought to investigate the feasibility and efficacy of the Glider balloon (GB), a side branch (SB) dedicated balloon 4-mm in length, in coronary bifurcation stenting.

METHODS AND RESULTS - In bifurcation bench models, stent configuration was examined with micro-focus computed tomography after crossover stenting followed by GB dilation or kissing balloon inflation (KBI). GB dilation maintained cross-sectional stent area without significant deformation and presented effective jailed strut removal in a high-angled bifurcation model.


We performed GB dilatation after main vessel (MV) stenting for 207 lesions in 194 patients, which included left main, true-bifurcation lesion, and two-stent treatment in 42.0%, 45.9%, and 14.0%, respectively. Proximal optimization technique (POT) or POT-like inflation was performed in 82.1%. GB crossing failure, SB stenting due to dissection, and stent deformation requiring correction by KBI or MV dilation occurred in 8.7, 1.4, and 5.8%, respectively. Finally, simple GB dilation without KBI had been completed in 91.8% for SB dilation. At 1-year follow-up, target lesion revascularization, cardiac death, myocardial infarction, and stent thrombosis were found in 7.2%, 2.1%, 2.1%, and 1.0%, respectively.


CONCLUSION - Simple GB dilation after adequate expansion of proximal MV stent provided acceptable acute and long-term results as an alternative to KBI.