CBS 2019
CBSMD教育中心
中 文

Bifurcation Stenting

Abstract

Recommended Article

One Versus 2-stent Strategy for the Treatment of Bifurcation Lesions in the Context of a Coronary Chronic Total Occlusion:A Multicenter Registry Comparison of intravascular ultrasound-guided with angiography-guided double kissing crush stenting for patients with complex coronary bifurcation lesions: rationale and design of a prospective, randomized and multicenter DKCRUSH VIII trial Streamlined reverse wire technique for the treatment of complex bifurcated lesions Percutaneous Coronary Intervention For Bifurcation Coronary Lesions.The 15th Consensus Document from the European Bifurcation Club Percutaneous Coronary Intervention Techniques for Bifurcation Disease: Network Meta-analysis Reveals Superiority of Double-Kissing Crush Technical aspects of the culotte technique Developing a Mobile Application for Global Cardiovascular Education Tips of the dual-lumen microcatheter-facilitated reverse wire technique in percutaneous coronary interventions for markedly angulated bifurcated lesions

Original ResearchAvailable online 10 September 2020

JOURNAL:Cardiovasc Revasc Med. Article Link

T and small protrusion (TAP) vs double kissing crush technique: Insights from in-vitro models

V Paradies, Jaryl Ng, HY Ang et al. Keywords: bifurcation lesions; double stenting; in-vitro models; thrombogenicity

ABSTRACT


BACKGROUND - Percutaneous coronary interventions on complex bifurcation lesions may require implantation of two stents to appropriately treat diffuse side-branch (SB) disease. Comparisons among different bifurcation stenting techniques are continuously attempted by various study designs (bench tests, computer simulations, clinical studies). Among different techniques, double kissing crush (DKC) represents the last evolution for crushingwhile T and small Protrusion (TAP) represents the evolution of T stenting. Both techniques are actually gaining popularity, but head-to-head comparisons are lacking.

 

METHODS AND RESULTS - Two last generation drug-eluting stents (Synergy, Boston Scientific, MA, USA and Ultimaster,Terumo Corp., Japan) were implanted in left main bifurcation bench models using TAP (n = 6 sets) and DKC (n = 6 sets) techniques. A peristaltic pump with fresh porcine blood was used to perfuse the blood through the silicone model at a flow rate of 200 ml/min for 4 min. Optical coherence tomography (OCT) was used to assess stent struts geometry and thrombus formation. SB cross sectional area as well as SB obstruction did not significantly differ between the two techniques. Numerical (but not statistically significant) differences were found in terms of malapposed struts (fewer with TAP) and floating struts (fewer with DKC). Thrombus formation after blood perfusion was similar between TAP and DKC technique (1.53 ± 1.12 vs. 1.20 ± 1.01 mm2, p = 0.6).

 

CONCLUSION - The result of the present in-vitro study shows the absence of significant difference between TAP and DKC in terms of stent struts apposition and acute thrombus formation potential. Despite the completely different technical steps required, both techniques have similar performance according to such articulated pre-clinical evaluation.