Outcomes 2 Years After Transcatheter Aortic Valve Replacement in Patients at Low Surgical Risk Prior Balloon Valvuloplasty Versus Direct Transcatheter Aortic Valve Replacement: Results From the DIRECTAVI Trial Suture- or Plug-Based Large-Bore Arteriotomy Closure: A Pilot Randomized Controlled Trial Procedural and clinical outcomes of type 0 versus type 1 bicuspid aortic valve stenosis undergoing trans-catheter valve replacement with new generation devices: Insight from the BEAT international collaborative registry Evaluation and Management of Aortic Stenosis in Chronic Kidney Disease: A Scientific Statement From the American Heart Association Transcatheter Aortic-Valve Replacement with a Balloon-Expandable Valve in Low-Risk Patients Aspirin with or without Clopidogrel after Transcatheter Aortic-Valve Implantation Clinical Impact of Valvular Heart Disease in Elderly Patients Admitted for Acute Coronary Syndrome: Insights From the Elderly-ACS 2 Study

Clinical Trial2020 Dec 16;S1936-8798(20)32011-2.

JOURNAL:JACC Cardiovasc Interv. Article Link

Suture- or Plug-Based Large-Bore Arteriotomy Closure: A Pilot Randomized Controlled Trial

MP van Wiechen, D Tchétché, N Dumonteil et al. Keywords: TAVR; vascular closure device; dedicated plug-based VCD vs suture-based VCD

ABSTRACT


OBJECTIVES - The authors sought to test the superiority in terms of efficacy and safety of a dedicated plug-based vascular closure device (VCD) during transcatheter aortic valve replacement (TAVR) over a suture-based VCD.


BACKGROUND - Vascular complications after TAVR are relevant and often associated with VCD failure.


METHODS - The MASH trial (MANTA vs. Suture-based vascular closure after transcatHeter aortic valve replacement) is an international, 2-center pilot randomized controlled trial comparing the MANTA VCD (Teleflex, Wayne, Pennsylvania) versus 2 ProGlides (Abbott Vascular, Abbott Park, Illinois). The primary composite endpoint consisted of access siterelated major or minor vascular complications at 30-daysfollow-up. Secondary endpoints included clinically relevant access site bleeding, time to hemostasis, and modified VCD failure (defined as failure to achieve hemostasis within 5 min or requiring additional endovascular maneuvers such as endovascular stenting, surgical techniques, or additional closure devices). Adverse events were adjudicated by an independent clinical events committee according to the VARC-2 definitions.


RESULTS - A total of 210 TAVR patients were included between October 2018 and January 2020. Median age was 81 years, 54% were male, and the median STS score was 2.7%. There was no significant difference in the primary endpoint of access siterelated vascular complications between MANTA and ProGlide (10% vs. 4%; p = 0.16). Clinically significant access site bleedings were similar with both closure techniques (9% vs. 6%; p = 0.57). Modified VCD failure occurred less frequently in MANTA versus ProGlide (20% vs. 40%; p < 0.01). Suture-based closure required more often additional closure devices, whereas MANTA numerically needed more covered stents and surgical bailouts.