CBS 2019
CBSMD教育中心
English

血管内超声指导

科研文章

荐读文献

Is intravascular ultrasound beneficial for percutaneous coronary intervention of bifurcation lesions? Evidence from a 4,314-patient registry Utility of intravascular ultrasound guidance in patients undergoing percutaneous coronary intervention for type C lesions Impact of Intravascular Ultrasound-Guided Drug-Eluting Stent Implantation on Patients With Chronic Kidney Disease: Subgroup Analysis From ULTIMATE Trial Comparison of inhospital mortality, length of hospitalization, costs, and vascular complications of percutaneous coronary interventions guided by ultrasound versus angiography Serial intravascular ultrasound assessment of very late stent thrombosis after sirolimus-eluting stent placement Coronary plaque redistribution after stent implantation is determined by lipid composition: A NIRS-IVUS analysis Relation between baseline plaque features and subsequent coronary artery remodeling determined by optical coherence tomography and intravascular ultrasound Randomized comparison of clinical outcomes between intravascular ultrasound and angiography-guided drug-eluting stent implantation for long coronary artery stenoses Effects of Intravascular Ultrasound-Guided Versus Angiography-Guided New-Generation Drug-Eluting Stent Implantation: Meta-Analysis With Individual Patient-Level Data From 2,345 Randomized Patients Comprehensive intravascular ultrasound assessment of stent area and its impact on restenosis and adverse cardiac events in 403 patients with unprotected left main disease

Original Research2009 Aug 1;74(2):348-56.

JOURNAL:Catheter Cardiovasc Interv. Article Link

Defining a new standard for IVUS optimized drug eluting stent implantation: the PRAVIO study

Gerber RT, Latib A, Colombo A et al. Keywords: IVUS guided PCI; angiography-guided PCI; DES; outcome

ABSTRACT


OBJECTIVES - Preliminary Investigation to the Angiographic Versus IVUS Optimization Trial is a single center prospective observational intravascular ultrasound (IVUS) guided stent implantation study assessing new criteria for optimal drug eluting stent (DES) deployment.


BACKGROUND - IVUS assessment of DES often reveals underexpansion and malapposition. Optimal stent deployment is currently poorly defined and previous criteria may not be suitable in long and complex lesions.


METHODS - Optimization was defined as achieving >/or 70% of the cross-sectional area (CSA) of the postdilation balloon. This criterion was applied in 113 complex lesions. The size of this balloon was calculated according to vessel media-to-media diameters at various sites inside the stented segment. The IVUS guided treated lesions were matched according to diabetes, vessel type, reference vessel diameter, minimum lumen diameter (MLD), and lesion length with a group of angiographic treated lesions to compare final MLD achieved.


RESULTS - Mean minimum stent CSA according to the postdilation balloon utilized was 4.62 mm(2), 6.26 mm(2), 7.87 mm(2), and 9.87 mm(2) for 2.5 mm, 3.0 mm, 3.5 mm, and 4 mm balloons, respectively. Final MLD (mm) was significantly larger in the IVUS compared to the angiographic-guided group (3.09 +/- 0.50 vs. 2.67 +/- 0.54; P < 0.0001). There were no procedural complications related to IVUS use.


CONCLUSIONS - We propose new IVUS criteria based on vessel remodeling that results in an increment in the final MLD, compared to angiographic guidance, which is much larger than any previously published study. This criterion seems to be safely achievable. A proposed randomized study (angiographic vs. IVUS optimization trial) has been launched to test these concepts.

 

(c) 2008 Wiley-Liss, Inc.