The conductive function of biopolymer corrects myocardial scar conduction blockage and resynchronizes contraction to prevent heart failure Chimney technique in a TAVR-in-TAVR procedure with high risk of left main artery ostium occlusion Outcomes 2 Years After Transcatheter Aortic Valve Replacement in Patients at Low Surgical Risk Dapagliflozin for treating chronic heart failure with reduced ejection fraction 2020 ACC/AHA Guideline for the Management of Patients With Valvular Heart Disease: A Report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines Anticoagulation in Concomitant Chronic Kidney Disease and Atrial Fibrillation: JACC Review Topic of the Week Cardiac Structural Changes After Transcatheter Aortic Valve Replacement: Systematic Review and Meta-Analysis of Cardiovascular Magnetic Resonance Studies Switching of Oral Anticoagulation Therapy After PCI in Patients With Atrial Fibrillation: The RE-DUAL PCI Trial Subanalysis Impact of epicardial adipose tissue on cardiovascular haemodynamics, metabolic profile, and prognosis in heart failure Rivaroxaban Plus Aspirin Versus Aspirin in Relation to Vascular Risk in the COMPASS Trial

Clinical TrialVolume 13, Issue 1, January 2020

JOURNAL:JACC Cardiovasc Interv. Article Link

Randomized Comparison Between Everolimus-Eluting Bioresorbable Scaffold and Metallic Stent: Multimodality Imaging Through 3 Years

Y Onuma, Y Honda, T Asano et al.

ABSTRACT


OBJECTIVES - The aim of this study was to investigate the vascular responses and fates of the scaffold after bioresorbable vascular scaffold (BVS) implantation using multimodality imaging.


BACKGROUND - Serial comprehensive image assessments after BVS implantation in the context of a randomized trial have not yet been reported.


METHODS - In the ABSORB Japan trial, 400 patients were randomized to a BVS (n = 266) or a cobalt-chromium everolimus-eluting stent (n = 134). Through 3 years, patients underwent serial angiography and intravascular ultrasound or optical coherence tomography (OCT).


RESULTS - Luminal dimension at 3 years was consistently smaller with the BVS than with the cobalt-chromium everolimus-eluting stent (mean angiographic minimal luminal diameter 2.04 ± 0.63 mm vs. 2.40 ± 0.56 mm, mean difference 0.37 mm [95% confidence interval: 0.50 to 0.24 mm]; p < 0.001), mainly because of smaller device area (6.13 ± 2.03 mm2 vs. 7.15 ± 2.16 mm2, mean difference 1.04 mm2 [95% confidence interval: 1.66 to 0.42 mm2]; p < 0.001), and larger neointimal area (2.10 ± 0.61 mm2 vs. 1.86 ± 0.64 mm2, mean difference 0.24 mm2 [95% confidence interval: 0.06 to 0.43 mm2]; p = 0.01) by OCT. BVS-treated vessels did not show previously reported favorable vessel responses, such as positive vessel remodeling, late luminal enlargement, and restoration of vasomotion, although the OCT-based healing score was on average zero (interquartile range: 0.00 to 0.00). At 3 years, intraluminal scaffold dismantling (ISD) was observed in 14% of BVS. On serial OCT, ISD was observed in 6 lesions at 2 years, where the struts had been fully apposed at post-procedure, while ISD was observed in 12 lesions at 3 years, where 8 lesions were free from ISD on 2-year OCT. In 5 cases of very late scaffold thrombosis, strut discontinuities were detected in all 4 cases with available OCT immediately before reintervention.


CONCLUSIONS- In this multimodality serial imaging study, luminal dimension at 3 years was smaller with the BVS than with the cobalt-chromium everolimus-eluting stent. ISD, suspected to be one of the mechanisms of very late BVS thrombosis, was observed in a substantial proportion of cases at 3 years, which developed between post-procedure and 2 years and even beyond 2 years. (AVJ-301 Clinical Trial: A Clinical Evaluation of AVJ-301 [AbsorbBVS] in Japanese Population [ABSORB JAPAN]; NCT01844284)