CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

Hemodynamic, Functional, and Clinical Responses to Pulmonary Artery Denervation in Patients With Pulmonary Arterial Hypertension of Different Causes Rivaroxaban for Thromboprophylaxis in High-Risk Ambulatory Patients With Cancer Long-Term Outcomes of Patients With Mediastinal Radiation–Associated Coronary Artery Disease Undergoing Coronary Revascularization With Percutaneous Coronary Intervention and Coronary Artery Bypass Grafting Systematic Review and Network Meta‐Analysis Comparing Bifurcation Techniques for Percutaneous Coronary Intervention Implications of the local hemodynamic forces on the formation and destabilization of neoatherosclerotic lesions Transseptal puncture versus patent foramen ovale or atrial septal defect access for left atrial appendage closure 2015 ACC/HRS/SCAI Left Atrial Appendage Occlusion Device Societal Overview Management and outcomes of patients with left atrial appendage thrombus prior to percutaneous closure Evolving insights into the role of local shear stress in late stent failure from neoatherosclerosis formation and plaque destabilization Alcohol consumption, cardiac biomarkers, and risk of atrial fibrillation and adverse outcomes

Clinical TrialVolume 13, Issue 1, January 2020

JOURNAL:JACC Cardiovasc Interv. Article Link

Randomized Comparison Between Everolimus-Eluting Bioresorbable Scaffold and Metallic Stent: Multimodality Imaging Through 3 Years

Y Onuma, Y Honda, T Asano et al.

ABSTRACT


OBJECTIVES - The aim of this study was to investigate the vascular responses and fates of the scaffold after bioresorbable vascular scaffold (BVS) implantation using multimodality imaging.


BACKGROUND - Serial comprehensive image assessments after BVS implantation in the context of a randomized trial have not yet been reported.


METHODS - In the ABSORB Japan trial, 400 patients were randomized to a BVS (n = 266) or a cobalt-chromium everolimus-eluting stent (n = 134). Through 3 years, patients underwent serial angiography and intravascular ultrasound or optical coherence tomography (OCT).


RESULTS - Luminal dimension at 3 years was consistently smaller with the BVS than with the cobalt-chromium everolimus-eluting stent (mean angiographic minimal luminal diameter 2.04 ± 0.63 mm vs. 2.40 ± 0.56 mm, mean difference 0.37 mm [95% confidence interval: 0.50 to 0.24 mm]; p < 0.001), mainly because of smaller device area (6.13 ± 2.03 mm2 vs. 7.15 ± 2.16 mm2, mean difference 1.04 mm2 [95% confidence interval: 1.66 to 0.42 mm2]; p < 0.001), and larger neointimal area (2.10 ± 0.61 mm2 vs. 1.86 ± 0.64 mm2, mean difference 0.24 mm2 [95% confidence interval: 0.06 to 0.43 mm2]; p = 0.01) by OCT. BVS-treated vessels did not show previously reported favorable vessel responses, such as positive vessel remodeling, late luminal enlargement, and restoration of vasomotion, although the OCT-based healing score was on average zero (interquartile range: 0.00 to 0.00). At 3 years, intraluminal scaffold dismantling (ISD) was observed in 14% of BVS. On serial OCT, ISD was observed in 6 lesions at 2 years, where the struts had been fully apposed at post-procedure, while ISD was observed in 12 lesions at 3 years, where 8 lesions were free from ISD on 2-year OCT. In 5 cases of very late scaffold thrombosis, strut discontinuities were detected in all 4 cases with available OCT immediately before reintervention.


CONCLUSIONS- In this multimodality serial imaging study, luminal dimension at 3 years was smaller with the BVS than with the cobalt-chromium everolimus-eluting stent. ISD, suspected to be one of the mechanisms of very late BVS thrombosis, was observed in a substantial proportion of cases at 3 years, which developed between post-procedure and 2 years and even beyond 2 years. (AVJ-301 Clinical Trial: A Clinical Evaluation of AVJ-301 [AbsorbBVS] in Japanese Population [ABSORB JAPAN]; NCT01844284)