Use of the Instantaneous Wave-free Ratio or Fractional Flow Reserve in PCI Instantaneous Wave-free Ratio versus Fractional Flow Reserve to Guide PCI Pulmonary arterial hypertension in congenital heart disease: an epidemiologic perspective from a Dutch registry Validation of bifurcation DEFINITION criteria and comparison of stenting strategies in true left main bifurcation lesions The Relation Between Optical Coherence Tomography-Detected Layered Pattern and Acute Side Branch Occlusion After Provisional Stenting of Coronary Bifurcation Lesions Comprehensive Management of Cardiovascular Risk Factors for Adults With Type 2 Diabetes: A Scientific Statement From the American Heart Association Cardiovascular risk prediction in type 2 diabetes: a comparison of 22 risk scores in primary care settings High-Resolution Cardiac Magnetic Resonance Imaging Techniques for the Identification of Coronary Microvascular Dysfunction Physiologic Characteristics and Clinical Outcomes of Patients With Discordance Between FFR and iFR Patient and Hospital Characteristics of Mitral Valve Surgery in the United States

Original ResearchVolume 13, Issue 6, March 2020

JOURNAL:JACC Cardiovasc Interv. Article Link

Feasibility of Coronary Access and Aortic Valve Reintervention in Low-Risk TAVR Patients

T Rogers, BC Greenspun, G Weissman et al. Keywords: coronary access; coronary artery obstruction; PCI; TAVR; valve-in-valve

ABSTRACT


OBJECTIVES - The aim of this study was to evaluate the feasibility of coronary access and aortic valve reintervention in low-risk patients undergoing transcatheter aortic valve replacement (TAVR) with a balloon-expandable transcatheter heart valve (THV).

 

BACKGROUND - Younger, low-risk TAVR patients are more likely than older, higher risk patients to require coronary angiography, percutaneous coronary intervention, or aortic valve reintervention, but their THVs may impede coronary access and cause coronary obstruction during TAVR-in-TAVR.

 

METHODS - The LRT (Low Risk TAVR) trial (NCT02628899) enrolled 200 subjects with symptomatic severe aortic stenosis to undergo TAVR using commercially available THVs. Subjects who received balloon-expandable THVs and who had 30-day cardiac computed tomographic scans were included in this study. In a subgroup, the feasibility of intentional THV crimping on the delivery catheter to pre-determine commissural alignment was tested.

 

RESULTS - In the LRT trial, 168 subjects received balloon-expandable THVs and had 30-day cardiac computed tomographic scans, of which 137 were of adequate image quality for analysis. The most challenging anatomy for coronary access (THV frame above and commissural suture post in front of a coronary ostium) was observed in 9% to 13% of subjects. Intentional THV crimping did not appear to meaningfully affect commissural alignment. The THV frame extended above the sinotubular junction in 21% of subjects, and in 13%, the distance between the THV and the sinotubular junction was <2 mm, signifying that TAVR-in-TAVR may not be feasible without causing coronary obstruction.

 