Prior Balloon Valvuloplasty Versus Direct Transcatheter Aortic Valve Replacement: Results From the DIRECTAVI Trial Regional Heterogeneity in the Coronary Vascular Response in Women With Chest Pain and Nonobstructive Coronary Artery Disease Longitudinal Assessment of Vascular Function With Sunitinib in Patients With Metastatic Renal Cell Carcinoma Intravascular Ultrasound and Angioscopy Assessment of Coronary Plaque Components in Chronic Totally Occluded Lesions Use of clopidogrel with or without aspirin in patients taking oral anticoagulant therapy and undergoing percutaneous coronary intervention: an open-label, randomised, controlled trial Exercise Intolerance in Patients With Heart Failure: JACC State-of-the-Art Review Bioprosthetic valve oversizing is associated with increased risk of valve thrombosis following TAVR Rationale and design of a large-scale, app-based study to identify cardiac arrhythmias using a smartwatch: The Apple Heart Study Predictors and Clinical Outcomes of Next-Day Discharge After Minimalist Transfemoral Transcatheter Aortic Valve Replacement Heart Failure With Recovered Left Ventricular Ejection Fraction: JACC Scientific Expert Panel

Original Research2020 Jul 5;S0167-5273(20)33411-2.

JOURNAL:Int J Cardiol . Article Link

Long-term outcome of prosthesis-patient mismatch after transcatheter aortic valve replacement

M Compagnone, G Marchetti, F Saia et al. Keywords: aortic valve stenosis; degenerated aortic valve bioprosthesis; prosthesis-patient mismatch; TAVR

ABSTRACT


BACKGROUND - Incidence and long-term clinical consequences of prosthesis-patient mismatch (PPM) after transcatheter aortic valve replacement (TAVR) are still unclear.


METHODS - We enrolled 710 consecutive patients who underwent TAVR. PPM was defined as absent if the index orifice area (iEOA) was >0.85 cm2/m2, moderate if the iEOA was between 0.65 and 0.85 cm2/m2 or severe if the iEOA was <0.65 cm2/m2.

RESULTS - Among the 566 patients fulfilling the study criteria, the distribution of PPM was as follows: 50.5% none ( n = 286), 43% moderate PPM ( n = 243) and 6.5% severe PPM ( n = 37). At 5-year follow-up, patients with severe PPM had a significantly higher incidence of the combined endpoint of cardiovascular death, acute myocardial infarction and stroke ( p = .025) compared with the other patients. After adjusting the results for possible confounders, severe PPM remained an independent predictor of long-term adverse outcome (HR: 2.46; 95% Confidence Interval: 1.10–5.53). The independent predictors of severe PPM were valve-in-valve procedure and body mass index. Balloon-expandable valves were not associated with higher rates of severe PPM in comparison with self-expandable valves (5% vs. 8%, respectively, p = .245).


CONCLUSIONS -In our study severe PPM emerged as a risk factor for long-term major adverse cardiac and cerebrovascular events.