Percutaneous coronary intervention using a combination of robotics and telecommunications by an operator in a separate physical location from the patient: an early exploration into the feasibility of telestenting (the REMOTE-PCI study) Coronary Artery Calcium Is Associated with Left Ventricular Diastolic Function Independent of Myocardial Ischemia 2-Year Outcomes After Stenting of Lipid-Rich and Nonrich Coronary Plaques Radionuclide Image-Guided Repair of the Heart A randomized multicentre trial to compare revascularization with optimal medical therapy for the treatment of chronic total coronary occlusions Macrophage MST1/2 Disruption Impairs Post-Infarction Cardiac Repair via LTB4 稳定性冠心病诊断与治疗指南 Management of Patients With NSTE-ACS: A Comparison of the Recent AHA/ACC and ESC Guidelines Qualitative Methodology in Cardiovascular Outcomes Research: A Contemporary Look Current Perspectives on Coronavirus Disease 2019 and Cardiovascular Disease: A White Paper by the JAHA Editors

Original Research2017 Jan 20;12(13):1569-1576.

JOURNAL:EuroIntervention. Article Link

Percutaneous coronary intervention using a combination of robotics and telecommunications by an operator in a separate physical location from the patient: an early exploration into the feasibility of telestenting (the REMOTE-PCI study)

Madder RD, VanOosterhout SM, Jacoby ME et al. Keywords: telestenting; robotic PCI; feasibility

ABSTRACT


AIMSThe present study explores the feasibility of telestenting, wherein a physician operator performs stenting on a patient in a separate physical location using a combination of robotics and telecommunications.


METHODS AND RESULTSPatients undergoing robotic stenting were eligible for inclusion. All manipulations of guidewires, balloons, and stents were performed robotically by a physician operator located in an isolated separate room outside the procedure room housing the patient. Communication between the operating physician and laboratory personnel was via telecommunication devices providing real-time audio and video connectivity. Among 20 patients who consented to participate, technical success, defined as successful advancement and retraction of guidewires, balloons, and stents by the robotic system without conversion to manual operation, was achieved in 19 of 22 lesions (86.4%). Procedural success, defined as <30% residual stenosis upon completion of the procedure in the absence of death or repeat revascularisation prior to hospital discharge, was achieved in 19 of 20 patients (95.0%). There were no deaths or repeat revascularisations prior to hospital discharge.

CONCLUSIONSTo the best of our knowledge, the present study is the first to explore the feasibility of telestenting. Additional studies are required to determine if future advancements in robotics will facilitate telestenting over greater geographic distances.