Oxidative Stress and Cardiovascular Risk: Obesity, Diabetes, Smoking, and Pollution: Part 3 of a 3-Part Series ACC Clinical Bulletin Focuses on Cardiac Implications of Coronavirus (COVID-19) CSC Expert Consensus on Principles of Clinical Management of Patients with Severe Emergent Cardiovascular Diseases during the COVID-19 Epidemic COVID-19 and Thrombotic or Thromboembolic Disease: Implications for Prevention, Antithrombotic Therapy, and Follow-up Antiinflammatory Therapy with Canakinumab for Atherosclerotic Disease Hemodynamic Response to Nitroprusside in Patients With Low-Gradient Severe Aortic Stenosis and Preserved Ejection Fraction Inflammation: A New Target For CAD Treatment and Prevention Percutaneous Repair or Medical Treatment for Secondary Mitral Regurgitation

Clinical TrialMarch 2020 Vol 13, Issue 3

JOURNAL:Circulation. Article Link

Randomized Comparison of Everolimus- and Zotarolimus-Eluting Coronary Stents With Biolimus-Eluting Stents in All-Comer Patients

YJ Youn, J-W Lee, J Yoon et al.

ABSTRACT


BACKGROUND - There is limited data comparing the Xience everolimus-eluting stent (EES) and the Resolute zotarolimus-eluting stent (ZES) with the BioMatrix biolimus-eluting stent (BES).


METHODS - This open-label, randomized, noninferiority trial enrolled all-comer patients to be randomly treated with either BES, EES, or ZES in a 1:1:1 ratio in 15 centers across South Korea. The primary end point was a device-oriented composite outcome consisting of cardiac death, target-vessel myocardial infarction, and clinically indicated target lesion revascularization at 24 months. The BES was compared with the EES and the ZES by intention-to-treat analyses with a noninferiority margin of 3.8%, respectively.


RESULTS - Because of slow recruitment and low event rates, this trial was prematurely terminated after enrollment of 1935 (75%) of the intended 2580 patients. Of the 1911 patients randomized to either EES (n=638), BES (n=634), or ZES (n =639), the rate of device-oriented composite outcome was 3.6%, 2.2%, and 3.9%, respectively, at 24 months (BES versus EES: absolute risk difference 1.4% [upper limit of 1-sided 95% CI: 3.2%]; Pfor noninferiority <0.001; BES versus ZES: absolute risk difference 1.7% [upper limit of 1-sided 95% CI: 3.6%]; Pfor noninferiority <0.001).


CONCLUSIONS - The BES was noninferior to either the EES or the ZES in all-comer patients for device-oriented composite outcome at the 24-month follow-up. However, caution is advised regarding interpretation of these results due to the premature termination of this study.