经导管主动脉瓣置换

Abstract

Recommended Article

Transcatheter Aortic Valve Replacement: Role of Multimodality Imaging in Common and Complex Clinical Scenarios Determinants and Impact of Heart Failure Readmission Following Transcatheter Aortic Valve Replacement Leaflet immobility and thrombosis in transcatheter aortic valve replacement Extracellular Myocardial Volume in Patients With Aortic Stenosis Randomized Evaluation of TriGuard 3 Cerebral Embolic Protection After Transcatheter Aortic Valve Replacement: REFLECT II Ambulatory Electrocardiogram Monitoring in Patients Undergoing Transcatheter Aortic Valve Replacement: JACC State-of-the-Art Review Ascending Aortic Length and Risk of Aortic Adverse Events: The Neglected Dimension Infective Endocarditis After Transcatheter Aortic Valve Replacement

Clinical Case Study2020 Nov 14;ehaa817.

JOURNAL:Eur Heart J. Article Link

Chimney technique in a TAVR-in-TAVR procedure with high risk of left main artery ostium occlusion

A Dibie, Q Landolff, A Veugeois et al. Keywords: clinical case; chimney technique; TAVR-in-TAVR procedure; left main artery ostium occlusion

ABSTRACT

An 87-year-old woman was referred to our centre for a valve-in-valve transcatheter aortic valve replacement (TAVR). She underwent a TAVR procedure with implantation of a SAPIEN XT 26 mm (Edwards Lifesciences) bioprosthesis 7 years before but subsequently developed symptomatic aortic valve failure (mean gradient = 64 mmHg).

Pre-procedural computed tomography (CT) scan identified a low height between the valvular plane and left main coronary artery (LMA) ostium equal to 2.7 mm (Panel A1). Valve simulation technique was applied: the virtual transcatheter heart valve/coronary distance was measured at 4.0 mm (Panel A2). The coronary angiography revealed no recurrent stenosis (Panel B)....