CBS 2019
CBSMD教育中心
中 文

Transcatheter Aortic Valve Replacement

Abstract

Recommended Article

Relationship Between Hospital Surgical Aortic Valve Replacement Volume and Transcatheter Aortic Valve Replacement Outcomes Coronary Access After TAVR Risk of Coronary Obstruction and Feasibility of Coronary Access After Repeat Transcatheter Aortic Valve Replacement With the Self-Expanding Evolut Valve: A Computed Tomography Simulation Study Transcatheter Aortic-Valve Replacement with a Self-Expanding Valve in Low-Risk Patients Transcatheter Versus Surgical Aortic Valve Replacement in Patients With Severe Aortic Valve Stenosis: 1-Year Results From the All-Comers NOTION Randomized Clinical Trial Informed Shared Decisions for Patients with Aortic Stenosis Transcatheter and surgical aortic valve replacement in patients with bicuspid aortic valve Coronary Protection to Prevent Coronary Obstruction During TAVR: A Multicenter International Registry

Original ResearchVolume 13, Issue 6, March 2020

JOURNAL:JACC Cardiovasc Interv. Article Link

Coronary Access After TAVR With a Self-Expanding Bioprosthesis: Insights From Computed Tomography

M Abdelghani, M Landt, H Traboulsi et al. Keywords: access; angiography; coronary intervention; transcatheter aortic valve

ABSTRACT


OBJECTIVES - The authors sought to estimate possible interference of the Medtronic Evolut R/Pro transcatheter heart valve (THV) frame with coronary access using multislice computed tomography (MSCT) data.

 

BACKGROUND - Lower-risk patients undergoing transcatheter aortic valve replacement (TAVR) endure a high cumulative risk of coronary events, but coronary access can be challenging.

 

METHODS - In 101 patients who received an Evolut R/Pro THV, post-TAVR MSCT (performed at a median of 30 days after TAVR) was used to assess possible interference of the elements of the THV frame with coronary access.

 

RESULTS - The closest cell of the THV frame vertically aligned with the coronary ostium was located opposite the ostium in 58% and 63%, below the ostium in 22% and 30%, or above the ostium in 20% and 7% of left and right coronary arteries, respectively. The free sinus of Valsalva space between the THV frame and the coronary ostium was 0.45 ± 0.17 cm and 0.44 ± 0.17 cm for the left and right coronary arteries, respectively, and showed a stepwise decrease with decreasing THV size (p < 0.001). Bioprosthetic valve commissures were antianatomic (i.e., not aligned with native commissures) in 45 patients (47%), and the commissural post was overlapping a coronary ostium in 15 patients (16%). Two patients (2.0%) had a possible interference of the paravalvular sealing skirt with coronary access.

 

CONCLUSIONS - Using post-TAVR MSCT data, the main mechanism of potential interference of Evolut R/Pro frame with coronary access was an antianatomic commissural post overlapping the coronary ostium.