CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

Reduced Apolipoprotein M and Adverse Outcomes Across the Spectrum of Human Heart Failure Transcatheter Aortic Valve Implantation Represents an Anti-Inflammatory Therapy Via Reduction of Shear Stress-Induced, Piezo-1-Mediated Monocyte Activation The Future of Biomarker-Guided Therapy for Heart Failure After the Guiding Evidence-Based Therapy Using Biomarker Intensified Treatment in Heart Failure (GUIDE-IT) Study Clinical Phenogroups in Heart Failure With Preserved Ejection Fraction: Detailed Phenotypes, Prognosis, and Response to Spironolactone Effect of SGLT2-Inhibitors on Epicardial Adipose Tissue: A Meta-Analysis Three vs twelve months of dual antiplatelet therapy after zotarolimus-eluting stents: the OPTIMIZE randomized trial Provisional versus elective two-stent strategy for unprotected true left main bifurcation lesions: Insights from a FAILS-2 sub-study ACC/AHA Versus ESC Guidelines on Dual Antiplatelet Therapy JACC Guideline Comparison: JACC State-of-the-Art Review The sinus venosus contributes to coronary vasculature through VEGFC-stimulated angiogenesis Noninvasive Imaging for the Evaluation of Diastolic Function: Promises Fulfilled

Original Research2014 Mar 20;9(11):1301-8.

JOURNAL:EuroIntervention. Article Link

Mechanical complications of everolimus-eluting stents associated with adverse events: an intravascular ultrasound study

Inaba S, Mintz GS, Yun KH et al. Keywords: intravascular ultrasound; everolimus-eluting stent; stent fracture

ABSTRACT


AIMS - Mechanical complications contribute to bare metal and first-generation drug-eluting stent (DES) failure. However, the importance of the mechanical complications of second-generation DES remains unclear. We report mechanical complications associated with everolimus-eluting stent (EES) failures.


METHODS AND RESULTS - We retrospectively analysed 177 consecutive EES-treated lesions in 136 patients who underwent intravascular ultrasound (IVUS) at follow-up. Mechanical complications were identified in 17 patients (five stable angina, 10 unstable angina, two non-ST-elevation myocardial infarction [NSTEMI] without angiographic thrombus). Fifteen (88.2%) were treated with repeat revascularisation. By IVUS, there were 16 focal (94.1%) and one diffuse (5.9%) in-stent restenoses. Complete stent fracture with separation was seen in only one, partial stent fracture with separation was seen in three, and in 13 there was longitudinal deformation (n=2) or stent strut fracture (n=11) with overlapping of the proximal and distal stent fragments. In 13 EES with evidence of overlapping in the setting of either fracture or deformation, there was a 35.5±12.2% smaller stent area compared to the adjacent proximal and distal stent fragments, and >50% neointimal hyperplasia in 12 (92.3%).

CONCLUSIONS - We found EES mechanical complications, often followed by longitudinal deformation or fracture leading to excessive neointimal hyperplasia, in-stent restenosis, and repeat revascularisation.