Pulmonary arterial hypertension in congenital heart disease: an epidemiologic perspective from a Dutch registry Ticagrelor versus clopidogrel in patients with acute coronary syndromes Thrombotic Versus Bleeding Risk After Transcatheter Aortic Valve Replacement: JACC Review Topic of the Week Ten-Year All-Cause Death According to Completeness of Revascularization in Patients With Three-Vessel Disease or Left Main Coronary Artery Disease: Insights From the SYNTAX Extended Survival Study Right ventricular function and outcome in patients undergoing transcatheter aortic valve replacement Intravascular Ultrasound Guidance vs. Angiographic Guidance in Primary Percutaneous Coronary Intervention for ST-Segment Elevation Myocardial Infarction - Long-Term Clinical Outcomes From the CREDO-Kyoto AMI Registry Regurgitant Volume/Left Ventricular End-Diastolic Volume Ratio: Prognostic Value in Patients With Secondary Mitral Regurgitation Inflammation and cholesterol as predictors of cardiovascular events among patients receiving statin therapy: a collaborative analysis of three randomised trials Correlations between fractional flow reserve and intravascular ultrasound in patients with an ambiguous left main coronary artery stenosis Anticoagulation with or without Clopidogrel after Transcatheter Aortic-Valve Implantation

Clinical Trial2016 Jun 13;9(11):1115-23.

JOURNAL:JACC Cardiovasc Interv. Article Link

Incidence and Clinical Outcomes of Stent Fractures on the Basis of 6,555 Patients and 16,482 Drug-Eluting Stents From 4 Centers

Kan J, Ge Z, Chen SL et al. Keywords: drug-eluting stent(s); restenosis; stent fracture; stent thrombosis; target lesion revascularization

ABSTRACT


OBJECTIVES - The present study aimed to analyze the incidence of SF and its correlation with clinical events after DES implantation and the outcome of re-intervention for symptomatic in-stent restenosis (ISR) induced by stent fracture (SF).


BACKGROUND - SF is associated with a high rate of clinical events after the implantation of drug-eluting stents (DES). However, the chronological rate of SF and the effect of SF on clinical outcomes from a large patient population remain underreported.

METHODS - A total of 6,555 patients with 16482 DES in 10751 diseased vessels and surveillance angiography between November 2003 and January 2014 were prospectively studied. The primary endpoints included the incidence of SF, in-stent restenosis (ISR), target lesion revascularization (TLR), and definite stent thrombosis (ST) at the end of follow-up before and after propensity score matching. Clinical outcomes after TLR were also followed up.

RESULTS - The SF rate was detected in 803 (12.3%) patients, 3,630 (22.0%) stents, and 1,852 (17.2%) diseased vessels. SF increased over time. SF was associated with higher unadjusted rates of ISR (42.1%), TLR (24.8%, n = 379), and definite ST (4.6%) compared with stents without fracture (10.7%, 6.6%, and 1.03%, all p < 0.001), and the differences remained significant after propensity score matching (all p < 0.05). There was no significant difference in any-cause or cardiac mortality between patients with and without SF. After 1,523 days of follow-up since the first surveillance angiography, repeat ISR was detected in 90 of 379 (23.8%) stents after reintervention, and 6 (7.5%) stents required repeat TLR.

CONCLUSIONS - SF is more frequently observed after DES implantation. TLR was required in almost one-fourth of fractured stents. Increased events in the SF group did not translate into a difference in mortality compared with the non-SF group. Reintervention was associated with acceptable clinical results.