CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

Benefit-risk profile of extended dual antiplatelet therapy beyond 1 year in patients with high risk of ischemic or bleeding events after PCI Safety of intermediate left main stenosis revascularization deferral based on fractional flow reserve and intravascular ultrasound: A systematic review and meta-regression including 908 deferred left main stenosis from 12 studies Prognostic Value of Computed Tomography-Derived Extracellular Volume in TAVR Patients With Low-Flow Low-Gradient Aortic Stenosis Empagliflozin, Health Status, and Quality of Life in Patients with Heart Failure and Preserved Ejection Fraction: The EMPEROR-Preserved Trial The Science Underlying COVID-19: Implications for the Cardiovascular System Stroke Complicating Infective Endocarditis After Transcatheter Aortic Valve Replacement Guideline Update on Indications for Transcatheter Aortic Valve Implantation Based on the 2020 American College of Cardiology/American Heart Association Guidelines for Management of Valvular Heart Disease Intravascular ultrasound in the evaluation and treatment of left main coronary artery disease: a consensus statement from the European Bifurcation Club Acute Aortic Syndrome Revisited: JACC State-of-the-Art Review A randomized comparison of Coronary Stents according to Short or Prolonged durations of Dual Antiplatelet Therapy in patients with Acute Coronary Syndromes: a pre-specified analysis of the SMART-DATE trial

Clinical Trial2016 Jun 13;9(11):1115-23.

JOURNAL:JACC Cardiovasc Interv. Article Link

Incidence and Clinical Outcomes of Stent Fractures on the Basis of 6,555 Patients and 16,482 Drug-Eluting Stents From 4 Centers

Kan J, Ge Z, Chen SL et al. Keywords: drug-eluting stent(s); restenosis; stent fracture; stent thrombosis; target lesion revascularization

ABSTRACT


OBJECTIVESThe present study aimed to analyze the incidence of SF and its correlation with clinical events after DES implantation and the outcome of re-intervention for symptomatic in-stent restenosis (ISR) induced by stent fracture (SF).


BACKGROUNDSF is associated with a high rate of clinical events after the implantation of drug-eluting stents (DES). However, the chronological rate of SF and the effect of SF on clinical outcomes from a large patient population remain underreported.

METHODSA total of 6,555 patients with 16482 DES in 10751 diseased vessels and surveillance angiography between November 2003 and January 2014 were prospectively studied. The primary endpoints included the incidence of SF, in-stent restenosis (ISR), target lesion revascularization (TLR), and definite stent thrombosis (ST) at the end of follow-up before and after propensity score matching. Clinical outcomes after TLR were also followed up.

RESULTSThe SF rate was detected in 803 (12.3%) patients, 3,630 (22.0%) stents, and 1,852 (17.2%) diseased vessels. SF increased over time. SF was associated with higher unadjusted rates of ISR (42.1%), TLR (24.8%, n = 379), and definite ST (4.6%) compared with stents without fracture (10.7%, 6.6%, and 1.03%, all p < 0.001), and the differences remained significant after propensity score matching (all p < 0.05). There was no significant difference in any-cause or cardiac mortality between patients with and without SF. After 1,523 days of follow-up since the first surveillance angiography, repeat ISR was detected in 90 of 379 (23.8%) stents after reintervention, and 6 (7.5%) stents required repeat TLR.

CONCLUSIONSSF is more frequently observed after DES implantation. TLR was required in almost one-fourth of fractured stents. Increased events in the SF group did not translate into a difference in mortality compared with the non-SF group. Reintervention was associated with acceptable clinical results.

Copyright © 2016 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.