Classic crush and DK crush stenting techniques Frequency of nonsystem delays in ST-elevation myocardial infarction patients undergoing primary percutaneous coronary intervention and implications for door-to-balloon time reporting (from the American Heart Association Mission: Lifeline program) The prognostic role of mid-range ejection fraction in ST-segment elevation myocardial infarction Radial Versus Femoral Access for Coronary Interventions Across the Entire Spectrum of Patients With Coronary Artery Disease: A Meta-Analysis of Randomized Trials Coronary Artery Plaque Characteristics Associated With Adverse Outcomes in the SCOT-HEART Study The HACD4 haplotype as a risk factor for atherosclerosis in males Changes in One-Year Mortality in Elderly Patients Admitted with Acute Myocardial Infarction in Relation with Early Management LOX-1 in Atherosclerosis and Myocardial Ischemia: Biology, Genetics, and Modulation The association between body mass index and obesity with survival in pulmonary arterial hypertension A sirolimus-eluting bioabsorbable polymer-coated stent (MiStent) versus an everolimus-eluting durable polymer stent (Xience) after percutaneous coronary intervention (DESSOLVE III): a randomised, single-blind, multicentre, non-inferiority, phase 3 trial

Clinical Trial2016 Jun 13;9(11):1115-23.

JOURNAL:JACC Cardiovasc Interv. Article Link

Incidence and Clinical Outcomes of Stent Fractures on the Basis of 6,555 Patients and 16,482 Drug-Eluting Stents From 4 Centers

Kan J, Ge Z, Chen SL et al. Keywords: drug-eluting stent(s); restenosis; stent fracture; stent thrombosis; target lesion revascularization

ABSTRACT


OBJECTIVESThe present study aimed to analyze the incidence of SF and its correlation with clinical events after DES implantation and the outcome of re-intervention for symptomatic in-stent restenosis (ISR) induced by stent fracture (SF).


BACKGROUNDSF is associated with a high rate of clinical events after the implantation of drug-eluting stents (DES). However, the chronological rate of SF and the effect of SF on clinical outcomes from a large patient population remain underreported.

METHODSA total of 6,555 patients with 16482 DES in 10751 diseased vessels and surveillance angiography between November 2003 and January 2014 were prospectively studied. The primary endpoints included the incidence of SF, in-stent restenosis (ISR), target lesion revascularization (TLR), and definite stent thrombosis (ST) at the end of follow-up before and after propensity score matching. Clinical outcomes after TLR were also followed up.

RESULTSThe SF rate was detected in 803 (12.3%) patients, 3,630 (22.0%) stents, and 1,852 (17.2%) diseased vessels. SF increased over time. SF was associated with higher unadjusted rates of ISR (42.1%), TLR (24.8%, n = 379), and definite ST (4.6%) compared with stents without fracture (10.7%, 6.6%, and 1.03%, all p < 0.001), and the differences remained significant after propensity score matching (all p < 0.05). There was no significant difference in any-cause or cardiac mortality between patients with and without SF. After 1,523 days of follow-up since the first surveillance angiography, repeat ISR was detected in 90 of 379 (23.8%) stents after reintervention, and 6 (7.5%) stents required repeat TLR.

CONCLUSIONSSF is more frequently observed after DES implantation. TLR was required in almost one-fourth of fractured stents. Increased events in the SF group did not translate into a difference in mortality compared with the non-SF group. Reintervention was associated with acceptable clinical results.

Copyright © 2016 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.