CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

Randomized study on simple versus complex stenting of coronary artery bifurcation lesions: the Nordic bifurcation study Contemporary techniques in percutaneous coronary intervention for bifurcation lesions Treatment of Very Small De Novo Coronary Artery Disease With 2.0 mm Drug-Coated Balloons Showed 1-Year Clinical Outcome Comparable With 2.0 mm Drug-Eluting Stents Influence of Local Myocardial Damage on Index of Microcirculatory Resistance and Fractional Flow Reserve in Target and Nontarget Vascular Territories in a Porcine Microvascular Injury Model Adaptive development of concomitant secondary mitral and tricuspid regurgitation after transcatheter aortic valve replacement Long-term efficacy and safety of drug-coated balloons versus drug-eluting stents for small coronary artery disease (BASKET-SMALL 2): 3-year follow-up of a randomised, non-inferiority trial Optical Coherence Tomography–Defined Plaque Vulnerability in Relation to Functional Stenosis Severity and Microvascular Dysfunction Clinical and angiographic outcomes of coronary dissection after paclitaxel-coated balloon angioplasty for small vessel coronary artery disease Drug-Coated Balloons for Coronary Artery Disease: Third Report of the International DCB Consensus Group Drug-Coated Balloons: A Safe and Effective Alternative to Drug-Eluting Stents in Small Vessel Coronary Artery Disease

Clinical TrialAugust 2018. [Epub ahead]

JOURNAL:JACC Cardiovasc Interv. Article Link

A Randomized Study of Distal Filter Protection Versus Conventional Treatment During Percutaneous Coronary Intervention in Patients With Attenuated Plaque Identified by Intravascular Ultrasound

K Hibi, K Kozuma, S Sonoda et al. Keywords: acute coronary syndrome(s); distal embolism; distal protection device; intravascular ultrasound

ABSTRACT


OBJECTIVES - The aim of this study was to evaluate the utility of distal protection during percutaneous coronary intervention (PCI) in patients with acute coronary syndromes at high risk for distal embolization.


BACKGROUND - The results of previous clinical trials indicated that the routine use of distal protection in patients with ST-segment elevation myocardial infarction did not improve clinical outcomes. However, selective use of distal protection by means of a filter-based distal protection system has not been evaluated.

METHODS - Two hundred patients with acute coronary syndromes who had native coronary artery lesions and attenuated plaque with longitudinal length ≥5 mm on pre-PCI intravascular ultrasound were randomly assigned to undergo PCI with distal protection or conventional treatment.

RESULTS - The primary endpoint (no-reflow phenomenon) occurred in 26 patients (26.5%) in the distal protection group and 39 patients (41.7%) in the conventional treatment group (p = 0.026), and the corrected TIMI (Thrombolysis In Myocardial Infarction) frame count after revascularization was significantly lower in the distal protection group (23 vs. 30.5; p = 0.0003). The incidence of cardiac death, cardiac arrest, cardiogenic shock after revascularization requiring defibrillation, cardiopulmonary resuscitation, or extracorporeal membrane oxygenation was significantly lower in the distal protection group than in the conventional treatment group (0% vs. 5.2%; p = 0.028).

CONCLUSIONS - The use of distal embolic protection applied with a filter device decreased the incidence of the no-reflow phenomenon and was associated with fewer serious adverse cardiac events after revascularization than conventional PCI in patients with acute coronary syndromes with attenuated plaque ≥5 mm in length. (Assessment of Distal Protection Device in Patients at High Risk for Distal Embolism in Acute Coronary Syndrome [ACS] [VAMPIRE3]; NCT01460966)