Optical coherence tomography is a kid on the block: I would choose intravascular ultrasound A systematic review of factors predicting door to balloon time in ST-segment elevation myocardial infarction treated with percutaneous intervention Correlation and prognostic role of neutrophil to lymphocyte ratio and SYNTAX score in patients with acute myocardial infarction treated with percutaneous coronary intervention: A six-year experience Biological Phenotypes of Heart Failure With Preserved Ejection Fraction Lower Risk of Heart Failure and Death in Patients Initiated on SGLT-2 Inhibitors Versus Other Glucose-Lowering Drugs: The CVD-REAL Study Outcomes in Patients Treated With Thin-Strut, Very Thin-Strut, or Ultrathin-Strut Drug-Eluting Stents in Small Coronary Vessels: A Prespecified Analysis of the Randomized BIO-RESORT Trial Pharmacoinvasive and Primary Percutaneous Coronary Intervention Strategies in ST-Elevation Myocardial Infarction (from the Mayo Clinic STEMI Network) Symptom onset-to-balloon time and mortality in the first seven years after STEMI treated with primary percutaneous coronary intervention Oxygen Therapy in Suspected Acute Myocardial Infarction HFpEF: From Mechanisms to Therapies

Original Research2018 Aug;31(4):421-429.

JOURNAL:J Interv Cardiol. Article Link

Impact of stent deformity induced by the kissing balloon technique for bifurcating lesions on in-stent restenosis after coronary intervention

Gamou T, Sataka K,Yamagishi M et al. Keywords: coronary bifurcation; intravascular ultrasound; kissing balloon technique

ABSRACT



OBJECTIVES - To investigate the impact of stent deformity induced by final kissing balloon technique (KBT) for coronary bifurcation lesions on in-stent restenosis (ISR).

 

BACKGROUND - In experimental models, the detrimental effects of KBT have been clearly demonstrated, but few data exists regarding the impact of proximal stent deformity induced by KBT on clinical outcomes.

 

METHODS - We examined 370 coronary lesions where intravascular ultrasound (IVUS)-guided second-generation drug-eluting stent (DES) implantation for coronary bifurcation lesions was performed. Based on IVUS analysis, the stent symmetry index (minimum/maximum stent diameter) and stent overstretch index (the mean of stent diameter/the mean of reference diameter) were calculated in the proximal main vessel.

 

RESULTS - The stent symmetry index was significantly lower (0.75 ± 0.07 vs 0.88 ± 0.06, P < 0.0001) and the stent overstretch index was significantly higher (1.04 ± 0.08 vs 1.01 ± 0.06, P = 0.0007) in lesions with KBT (n = 174) compared to those without KBT (n = 196). The number of two-stent technique in lesions with KBT was 31 (18%). In multivariate analysis, the degree of stent deformity indices was not associated with ISR in lesions with KBT; however, two-stent technique use was the only independent predictor of ISR at 8 months (hazard ratio: 3.96, 95% confidence interval: 1.25-12.5, P = 0.01).

 

CONCLUSIONS - Second-generation DES deformity induced by KBT was not associated with mid-term ISR.

 

 

© 2018 Wiley Periodicals, Inc.