Randomized Evaluation of TriGuard 3 Cerebral Embolic Protection After Transcatheter Aortic Valve Replacement: REFLECT II Extracellular Myocardial Volume in Patients With Aortic Stenosis Computed tomography angiography-derived extracellular volume fraction predicts early recovery of left ventricular systolic function after transcatheter aortic valve replacement Impact of Pre-Existing and New-Onset Atrial Fibrillation on Outcomes After Transcatheter Aortic Valve Replacement Long-term health outcome and mortality evaluation after invasive coronary treatment using drug eluting stents with or without the IVUS guidance. Randomized control trial. HOME DES IVUS Impact of Intravascular Ultrasound on Long-Term Clinical Outcomes in Patients With Acute Myocardial Infarction Thrombotic Versus Bleeding Risk After Transcatheter Aortic Valve Replacement: JACC Review Topic of the Week Contemporary Use and Trends in Unprotected Left Main Coronary Artery Percutaneous Coronary Intervention in the United States: An Analysis of the National Cardiovascular Data Registry Research to Practice Initiative Intravascular ultrasound-guided percutaneous coronary intervention improves the clinical outcome in patients undergoing multiple overlapping drug-eluting stents implantation Left Ventricular Rapid Pacing Via the Valve Delivery Guidewire in Transcatheter Aortic Valve Replacement

Clinical Trial2018 Aug 16.

JOURNAL:Cardiovasc Revasc Med. Article Link

A randomised trial comparing two stent sizing strategies in coronary bifurcation treatment with bioresorbable vascular scaffolds - The Absorb Bifurcation Coronary (ABC) trial

Rampat R, Mayo T, Hildick-Smith D et al. Keywords: Absorb; BVS; Bioresorbable vascular scaffold; Coronary bifurcation

ABSTRACT


BACKGROUNDLimited information is available on the use of Bioresorbable Vascular Scaffold (BVS) in bifurcations involving significant side branches. When treating bifurcation disease with metal stents, the recommendation is to choose a stent diameter based on the distal main vessel diameter. Whether this sizing strategy is applicable to BVS is currently unknown.


METHODS - We randomised 37 patients undergoing elective PCI for 'false' bifurcation disease (Medina 0,1,0; 1,0,0; 1,1,0) to receive BVS based either on proximal or distal reference diameters. Optical Frequency Domain Imaging (OFDI) measurements were performed pre BVS insertion to obtain proximal and distal reference diameters and post implantation. BVS size was chosen according to the proximal or distal reference diameter as per randomisation. Implantation was performed using the PSP technique tailored to bifurcation stenting. OFDI was repeated post implantation to confirm satisfactory expansion and apposition.

RESULTS - Baseline demographics between the two groups were similar. Patients were aged 62.8 ± 3.3 years; 76% were male. Mean side branch diameter was 2.24 ± 0.13 mm. TIMI III flow in the main vessel was achieved in all cases. Side branch occlusion occurred in 1 case (2.7%). In the distal-sizing arm, there was a greater incidence of significant malapposition (>300 μm) at the proximal end of the scaffold on OCT (2.3% versus 0.8%, p 0.023). The incidence of distal edge dissections was numerically greater in the proximal-sizing group but this was not statistically significant (31.3% vs 11.8%, p 0.17).

CONCLUSION - Both proximal and distal sizing strategies have similar procedural complication rates when using the ABSORB BVS to treat coronary bifurcations. However a proximal sizing strategy is associated with less malapposition and may be preferable.

Copyright © 2018 Elsevier Inc. All rights reserved.