Influence of LDL-Cholesterol Lowering on Cardiovascular Outcomes in Patients With Diabetes Mellitus Undergoing Coronary Revascularization 2-year outcomes with the Absorb bioresorbable scaffold for treatment of coronary artery disease: a systematic review and meta-analysis of seven randomised trials with an individual patient data substudy Incidence, Predictors, and Outcomes of In-Hospital Percutaneous Coronary Intervention Following Coronary Artery Bypass Grafting Impact of Optimal Medical Therapy on 10-Year Mortality After Coronary Revascularization Cardiovascular Risk Reduction with Icosapent Ethyl for Hypertriglyceridemia Rare Genetic Variants Associated With Sudden Cardiac Death in Adults Timing and Causes of Unplanned Readmissions After Percutaneous Coronary Intervention: Insights From the Nationwide Readmission Database Effect of a Home-Based Wearable Continuous ECG Monitoring Patch on Detection of Undiagnosed Atrial Fibrillation The mSToPS Randomized Clinical Trial Use of High-Risk Coronary Atherosclerotic Plaque Detection for Risk Stratification of Patients With Stable Chest Pain: A Secondary Analysis of the PROMISE Randomized Clinical Trial Defining High Bleeding Risk in Patients Undergoing Percutaneous Coronary Intervention: A Consensus Document From the Academic Research Consortium for High Bleeding Risk

Clinical Trial2018 Aug 16.

JOURNAL:Cardiovasc Revasc Med. Article Link

A randomised trial comparing two stent sizing strategies in coronary bifurcation treatment with bioresorbable vascular scaffolds - The Absorb Bifurcation Coronary (ABC) trial

Rampat R, Mayo T, Hildick-Smith D et al. Keywords: Absorb; BVS; Bioresorbable vascular scaffold; Coronary bifurcation

ABSTRACT


BACKGROUNDLimited information is available on the use of Bioresorbable Vascular Scaffold (BVS) in bifurcations involving significant side branches. When treating bifurcation disease with metal stents, the recommendation is to choose a stent diameter based on the distal main vessel diameter. Whether this sizing strategy is applicable to BVS is currently unknown.


METHODS - We randomised 37 patients undergoing elective PCI for 'false' bifurcation disease (Medina 0,1,0; 1,0,0; 1,1,0) to receive BVS based either on proximal or distal reference diameters. Optical Frequency Domain Imaging (OFDI) measurements were performed pre BVS insertion to obtain proximal and distal reference diameters and post implantation. BVS size was chosen according to the proximal or distal reference diameter as per randomisation. Implantation was performed using the PSP technique tailored to bifurcation stenting. OFDI was repeated post implantation to confirm satisfactory expansion and apposition.

RESULTS - Baseline demographics between the two groups were similar. Patients were aged 62.8 ± 3.3 years; 76% were male. Mean side branch diameter was 2.24 ± 0.13 mm. TIMI III flow in the main vessel was achieved in all cases. Side branch occlusion occurred in 1 case (2.7%). In the distal-sizing arm, there was a greater incidence of significant malapposition (>300 μm) at the proximal end of the scaffold on OCT (2.3% versus 0.8%, p 0.023). The incidence of distal edge dissections was numerically greater in the proximal-sizing group but this was not statistically significant (31.3% vs 11.8%, p 0.17).

CONCLUSION - Both proximal and distal sizing strategies have similar procedural complication rates when using the ABSORB BVS to treat coronary bifurcations. However a proximal sizing strategy is associated with less malapposition and may be preferable.