CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

Coronary Artery Calcium Is Associated with Left Ventricular Diastolic Function Independent of Myocardial Ischemia Advances in Coronary No-Reflow Phenomenon-a Contemporary Review A Randomized Trial Comparing the NeoVas Sirolimus-Eluting Bioresorbable Scaffold and Metallic Everolimus-Eluting Stents Homeostatic Chemokines and Prognosis in Patients With Acute Coronary Syndromes Better Prognosis After Complete Revascularization Using Contemporary Coronary Stents in Patients With Chronic Kidney Disease Long-Term Effect of Ultrathin-Strut Versus Thin-Strut Drug-Eluting Stents in Patients With Small Vessel Coronary Artery Disease Undergoing Percutaneous Coronary Intervention: A Subgroup Analysis of the BIOSCIENCE Randomized Trial Radionuclide Image-Guided Repair of the Heart Management of Patients With NSTE-ACS: A Comparison of the Recent AHA/ACC and ESC Guidelines Left Ventricular Assist Devices for Lifelong Support Percutaneous Support Devices for Percutaneous Coronary Intervention

Original Research2007 Aug 15;100(4):615-20.

JOURNAL:Am J Cardiol. Article Link

Intravascular Ultrasound Parameters Associated With Stent Thrombosis After Drug-Eluting Stent Deployment

Okabe T, Mintz GS, Buch AN et al. Keywords: DES; thrombosis; IVUS; in-stent restenosis

ABSTRACT


Drug-eluting stent (DES) thrombosis (ST) can be devastating. The study aim was to evaluate intravascular ultrasound (IVUS) predictors for DES thrombosis by comparing IVUS studies after implantation in 13 patients with 14 DES thrombosis lesions with a group of controls (30 lesions in 27 patients) matched for history of chronic renal failure and type of DES. Five patients (38%) discontinued dual antiplatelet therapy at the time of ST. There were 3 in-stent restenosis lesions (21%) treated using DESs in the ST group compared with 0 in the control group (p <0.05). Compared with the control group, IVUS studies in the ST group showed a smaller minimum stent area (4.6 +/- 1.1 vs 5.6 +/- 1.7 mm(2), p = 0.0489). In the ST group, 11 of 14 stents had a minimum stent area < or =5.0 mm(2) compared with 12 of 30 in the control group (p = 0.0392). Minimum stent area in patients who stopped clopidogrel therapy and developed ST (5.30 +/- 1.15 mm(2)) tended to be larger compared with that in patients who developed ST while using clopidogrel (4.24 +/- 0.96 mm(2), p = 0.091). Within the 5-mm-long proximal and distal reference segments analyzed, the ST group had larger proximal reference maximum plaque burdens and smaller minimum lumen areas, along with a tendency toward similar findings in the distal reference segments. In conclusion, IVUS findings at the time of DES implantation in patients who subsequently developed ST showed a smaller minimum stent area (especially in patients who developed ST while using clopidogrel) and more residual disease at the stent edges.