Is Acute heart failure a distinctive disorder? An analysis from BIOSTAT-CHF Prognostic implication of lipidomics in patients with coronary total occlusion undergoing PCI Left Ventricular Rapid Pacing Via the Valve Delivery Guidewire in Transcatheter Aortic Valve Implantation Significantly less inappropriate shocks in ischemic patients compared to non-ischemic patients: The S-ICD experience of a high volume single-center A Genotype-Guided Strategy for Oral P2Y12 Inhibitors in Primary PCI Incidence and Clinical Outcomes of Stent Fractures on the Basis of 6,555 Patients and 16,482 Drug-Eluting Stents From 4 Centers Intravascular ultrasound findings of early stent thrombosis after primary percutaneous intervention in acute myocardial infarction: a Harmonizing Outcomes with Revascularization and Stents in Acute Myocardial Infarction (HORIZONS-AMI) substudy 3D Printing and Heart Failure: The Present and the Future Coronary bifurcation lesions treated with simple or complex stenting: 5-year survival from patient-level pooled analysis of the Nordic Bifurcation Study and the British Bifurcation Coronary Study Blood Pressure Assessment in Adults in Clinical Practice and Clinic-Based Research: JACC Scientific Expert Panel

Original Research2007 Aug 15;100(4):615-20.

JOURNAL:Am J Cardiol. Article Link

Intravascular Ultrasound Parameters Associated With Stent Thrombosis After Drug-Eluting Stent Deployment

Okabe T, Mintz GS, Buch AN et al. Keywords: DES; thrombosis; IVUS; in-stent restenosis

ABSTRACT


Drug-eluting stent (DES) thrombosis (ST) can be devastating. The study aim was to evaluate intravascular ultrasound (IVUS) predictors for DES thrombosis by comparing IVUS studies after implantation in 13 patients with 14 DES thrombosis lesions with a group of controls (30 lesions in 27 patients) matched for history of chronic renal failure and type of DES. Five patients (38%) discontinued dual antiplatelet therapy at the time of ST. There were 3 in-stent restenosis lesions (21%) treated using DESs in the ST group compared with 0 in the control group (p <0.05). Compared with the control group, IVUS studies in the ST group showed a smaller minimum stent area (4.6 +/- 1.1 vs 5.6 +/- 1.7 mm(2), p = 0.0489). In the ST group, 11 of 14 stents had a minimum stent area < or =5.0 mm(2) compared with 12 of 30 in the control group (p = 0.0392). Minimum stent area in patients who stopped clopidogrel therapy and developed ST (5.30 +/- 1.15 mm(2)) tended to be larger compared with that in patients who developed ST while using clopidogrel (4.24 +/- 0.96 mm(2), p = 0.091). Within the 5-mm-long proximal and distal reference segments analyzed, the ST group had larger proximal reference maximum plaque burdens and smaller minimum lumen areas, along with a tendency toward similar findings in the distal reference segments. In conclusion, IVUS findings at the time of DES implantation in patients who subsequently developed ST showed a smaller minimum stent area (especially in patients who developed ST while using clopidogrel) and more residual disease at the stent edges.