CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

Optical coherence tomography is a kid on the block: I would choose intravascular ultrasound A systematic review of factors predicting door to balloon time in ST-segment elevation myocardial infarction treated with percutaneous intervention Correlation and prognostic role of neutrophil to lymphocyte ratio and SYNTAX score in patients with acute myocardial infarction treated with percutaneous coronary intervention: A six-year experience Biological Phenotypes of Heart Failure With Preserved Ejection Fraction Lower Risk of Heart Failure and Death in Patients Initiated on SGLT-2 Inhibitors Versus Other Glucose-Lowering Drugs: The CVD-REAL Study Outcomes in Patients Treated With Thin-Strut, Very Thin-Strut, or Ultrathin-Strut Drug-Eluting Stents in Small Coronary Vessels: A Prespecified Analysis of the Randomized BIO-RESORT Trial Pharmacoinvasive and Primary Percutaneous Coronary Intervention Strategies in ST-Elevation Myocardial Infarction (from the Mayo Clinic STEMI Network) Symptom onset-to-balloon time and mortality in the first seven years after STEMI treated with primary percutaneous coronary intervention Oxygen Therapy in Suspected Acute Myocardial Infarction HFpEF: From Mechanisms to Therapies

Original Research2007 Aug 15;100(4):615-20.

JOURNAL:Am J Cardiol. Article Link

Intravascular Ultrasound Parameters Associated With Stent Thrombosis After Drug-Eluting Stent Deployment

Okabe T, Mintz GS, Buch AN et al. Keywords: DES; thrombosis; IVUS; in-stent restenosis

ABSTRACT


Drug-eluting stent (DES) thrombosis (ST) can be devastating. The study aim was to evaluate intravascular ultrasound (IVUS) predictors for DES thrombosis by comparing IVUS studies after implantation in 13 patients with 14 DES thrombosis lesions with a group of controls (30 lesions in 27 patients) matched for history of chronic renal failure and type of DES. Five patients (38%) discontinued dual antiplatelet therapy at the time of ST. There were 3 in-stent restenosis lesions (21%) treated using DESs in the ST group compared with 0 in the control group (p <0.05). Compared with the control group, IVUS studies in the ST group showed a smaller minimum stent area (4.6 +/- 1.1 vs 5.6 +/- 1.7 mm(2), p = 0.0489). In the ST group, 11 of 14 stents had a minimum stent area < or =5.0 mm(2) compared with 12 of 30 in the control group (p = 0.0392). Minimum stent area in patients who stopped clopidogrel therapy and developed ST (5.30 +/- 1.15 mm(2)) tended to be larger compared with that in patients who developed ST while using clopidogrel (4.24 +/- 0.96 mm(2), p = 0.091). Within the 5-mm-long proximal and distal reference segments analyzed, the ST group had larger proximal reference maximum plaque burdens and smaller minimum lumen areas, along with a tendency toward similar findings in the distal reference segments. In conclusion, IVUS findings at the time of DES implantation in patients who subsequently developed ST showed a smaller minimum stent area (especially in patients who developed ST while using clopidogrel) and more residual disease at the stent edges.