CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

Sirolimus-eluting stent implantation for unprotected left main coronary artery stenosis: comparison with bare metal stent implantation Impact of chronic obstructive pulmonary disease on prognosis after percutaneous coronary intervention and bypass surgery for left main coronary artery disease: an analysis from the EXCEL trial Outcomes of procedural complications in transfemoral transcatheter aortic valve replacement Streamlined reverse wire technique for the treatment of complex bifurcated lesions Fate of post-procedural malapposition of everolimus-eluting polymeric bioresorbable scaffold and everolimus-eluting cobalt chromiummetallic stent in human coronary arteries: sequential assessment with optical coherence tomography in ABSORB Japan trial Superficial Calcium Fracture After PCI as Assessed by OCT Survival prospects of treatment naïve patients with Eisenmenger: a systematic review of the literature and report of own experience Sotatercept for the Treatment of Pulmonary Arterial Hypertension Changes in Coronary Plaque Composition in Patients With Acute Myocardial Infarction Treated With High-Intensity Statin Therapy (IBIS-4): A Serial Optical Coherence Tomography Study Long-Term Clinical Outcomes and Optimal Stent Strategy in Left Main Coronary Bifurcation Stenting

Original Research 2018 Sep 30. [Epub ahead of print]

JOURNAL:Catheter Cardiovasc Interv. Article Link

Novel predictor of target vessel revascularization after coronary stent implantation: Intraluminal intensity of blood speckle on intravascular ultrasound

Saito Y, Kitahara H, Okuya Y et al. Keywords: intravascular ultrasound; percutaneous coronary intervention; target vessel revascularization

ABSTRACT


BACKGROUND - The difference in intraluminal intensity of blood speckle (IBS) on integrated backscatter-intravascular ultrasound (IB-IVUS) across the coronary artery stenosis (i.e., ΔIBS) has been reported to negatively correlate with fractional flow reserve. Fractional flow reserve after coronary stenting is known as a predictor of target vessel revascularization (TVR). However, the relation between ΔIBS and TVR is unclear.


METHODS - Seven hundred and three vessels which underwent percutaneous coronary intervention with stents were screened. Vessels without IVUS-guidance and follow-up information were excluded. Intraluminal IBS values were measured using IB-IVUS in cross-sections at the ostium of the target vessel and at the distal reference of implanted stent. ΔIBS was calculated as (distal IBS) - (ostium IBS).


RESULTS - A total of 393 vessels were included. Mean ΔIBS at postprocedure was 6.22 ± 5.65. During the follow-up period (11.2 ± 3.1 months), 24 cases (6.1%) had TVR. ΔIBS was significantly greater in the vessels with TVR than in those without (11.10 ± 5.93 vs. 5.90 ± 5.49, P <0.001). In receiver operating characteristic curve analysis, ΔIBS significantly predicted TVR (AUC 0.74, best cut-off value 8.24, P < 0.001). Multiple logistic regression analysis showed use of drug eluting stent and ΔIBS ≥ 8.24 as independent predictors of TVR.


CONCLUSIONS - ΔIBS at postprocedure was significantly associated with TVR. IVUS may be able to predict TVR by physiological assessment with measurement of ΔIBS.


© 2018 Wiley Periodicals, Inc.