CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

Use of IVUS guided coronary stenting with drug eluting stent: a systematic review and meta-analysis of randomized controlled clinical trials and high quality observational studies Reduced Apolipoprotein M and Adverse Outcomes Across the Spectrum of Human Heart Failure Updated Expert Consensus Statement on Platelet Function and Genetic Testing for Guiding P2Y12 Receptor Inhibitor Treatment in Percutaneous Coronary Intervention rhACE2 Therapy Modifies Bleomycin-Induced Pulmonary Hypertension via Rescue of Vascular Remodeling Percutaneous Left Atrial Appendage Closure for Stroke Prophylaxis in Patients With Atrial Fibrillation: 2.3-Year Follow-up of the PROTECT AF (Watchman Left Atrial Appendage System for Embolic Protection in Patients With Atrial Fibrillation) Trial Transcatheter Aortic Valve Implantation Represents an Anti-Inflammatory Therapy Via Reduction of Shear Stress-Induced, Piezo-1-Mediated Monocyte Activation Three vs twelve months of dual antiplatelet therapy after zotarolimus-eluting stents: the OPTIMIZE randomized trial Management of Antithrombotic Therapy in Atrial Fibrillation Patients Undergoing PCI: JACC State-of-the-Art Review Clinical Phenogroups in Heart Failure With Preserved Ejection Fraction: Detailed Phenotypes, Prognosis, and Response to Spironolactone Provisional versus elective two-stent strategy for unprotected true left main bifurcation lesions: Insights from a FAILS-2 sub-study

Original Research 2018 Oct 23.

JOURNAL:Catheter Cardiovasc Interv. Article Link

In vitro flow and optical coherence tomography comparison of two bailout techniques after failed provisional stenting for bifurcation percutaneous coronary interventions

Adjedj J, Picard F, Mogi S et al. Keywords: Reversed String; T and protrusion; bifurcation angle; in vitro bifurcation models; provisional T-stenting

ABSTRACT


OBJECTIVES - To evaluate, in vitro, SB stenting techniques after failed provisional stenting. We aimed to compare flows and stent strut apposition of T and protrusion (TAP) versus Reversed String (RS) techniques using a flow simulator, optical coherence tomography (OCT) using silicon bifurcation phantoms with different bifurcation angulations.

 

BACKGROUND -  While bifurcation coronary artery stenoses are preferably treated with provisional T-stenting strategy, the preferred bailout two stents technique to treat the side branch remains unclear.

 

METHODS AND RESULTS - Eleven 30°-angle and ten 60°-angle bifurcatio nphantoms were used. After performing provisional stenting, TAP and RS techniques were compared in six phantoms with 30° and five with 60° angles. Flow measurement was performed using absolute coronary flow and particle image velocimetry techniques. Strut apposition was evaluated using OCT. Flow analyses showed that disturbed flow regions were observed in the vicinity of floating struts protruded into the lumen both regardless of TAP and RS techniques. OCT analysis showed a higher proportion of floating struts protruding into the main branch with TAP compared to RS, respectively (13% vs. 1%; P <0.001) in both angles.

 

CONCLUSIONS - RS reduces the proportion of floating struts protruding into the main branch compared to TAP, at comparable flow rates. Clinical studies are needed to evaluate feasibility and potential clinical benefit of this technique.

 

© 2018 Wiley Periodicals, Inc.