Two-year outcomes following unprotected left main stenting with first vs new-generation drug-eluting stents: the FINE registry. EuroIntervention. Genotyping to Guide Clopidogrel Treatment: An In-Depth Analysis of the TAILOR-PCI Trial P2Y12 Inhibitor Monotherapy with Clopidogrel Versus Ticagrelor in Patients with Acute Coronary Syndrome Undergoing Percutaneous Coronary Intervention Long-Term Outcomes After PCI or CABG for Left Main Coronary Artery Disease According to Lesion Location The Year in Cardiovascular Medicine 2020: Coronary Prevention: Looking back on the Year in Cardiovascular Medicine for 2020 in the field of coronary prevention is Professor Ramon Estruch, Dr Luis Ruilope, and Professor Francesco Cosentino. Mark Nicholls meets them Ticagrelor versus clopidogrel in patients with acute coronary syndromes Positive remodeling at 3 year follow up is associated with plaque-free coronary wall segment at baseline: a serial IVUS study Left Ventricular Hypertrophy and Clinical Outcomes Over 5 Years After TAVR: An Analysis of the PARTNER Trials and Registries Impact of Intravascular Ultrasound on Long-Term Clinical Outcomes in Patients With Acute Myocardial Infarction Mechanisms of in-stent restenosis after drug-eluting stent implantation: intravascular ultrasound analysis

Original Research 2018 Oct 23.

JOURNAL:Catheter Cardiovasc Interv. Article Link

In vitro flow and optical coherence tomography comparison of two bailout techniques after failed provisional stenting for bifurcation percutaneous coronary interventions

Adjedj J, Picard F, Mogi S et al. Keywords: Reversed String; T and protrusion; bifurcation angle; in vitro bifurcation models; provisional T-stenting

ABSTRACT


OBJECTIVES - To evaluate, in vitro, SB stenting techniques after failed provisional stenting. We aimed to compare flows and stent strut apposition of T and protrusion (TAP) versus Reversed String (RS) techniques using a flow simulator, optical coherence tomography (OCT) using silicon bifurcation phantoms with different bifurcation angulations.

 

BACKGROUND -  While bifurcation coronary artery stenoses are preferably treated with provisional T-stenting strategy, the preferred bailout two stents technique to treat the side branch remains unclear.

 

METHODS AND RESULTS - Eleven 30°-angle and ten 60°-angle bifurcatio nphantoms were used. After performing provisional stenting, TAP and RS techniques were compared in six phantoms with 30° and five with 60° angles. Flow measurement was performed using absolute coronary flow and particle image velocimetry techniques. Strut apposition was evaluated using OCT. Flow analyses showed that disturbed flow regions were observed in the vicinity of floating struts protruded into the lumen both regardless of TAP and RS techniques. OCT analysis showed a higher proportion of floating struts protruding into the main branch with TAP compared to RS, respectively (13% vs. 1%; P <0.001) in both angles.

 

CONCLUSIONS - RS reduces the proportion of floating struts protruding into the main branch compared to TAP, at comparable flow rates. Clinical studies are needed to evaluate feasibility and potential clinical benefit of this technique.

 

© 2018 Wiley Periodicals, Inc.