Left Ventricular Assist Devices for Lifelong Support Coronary flow velocity reserve predicts adverse prognosis in women with angina and noobstructive coronary artery disease: resultsfrom the iPOWER study Optimal Stenting Technique for Complex Coronary Lesions Intracoronary Imaging-Guided Pre-Dilation, Stent Sizing, and Post-Dilation Guiding Principles for Chronic Total Occlusion Percutaneous Coronary Intervention Statin Safety and Associated Adverse Events: A Scientific Statement From the American Heart Association Cardiovascular Biomarkers and Imaging in Older Adults: JACC Council Perspectives Association of Thrombus Aspiration With Time and Mortality Among Patients With ST-Segment Elevation Myocardial Infarction: A Post Hoc Analysis of the Randomized TOTAL Trial 2013 ACC/AHA Guideline on the Treatment of Blood Cholesterol to Reduce Atherosclerotic Cardiovascular Risk in Adults: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines 稳定性冠心病诊断与治疗指南 Ticagrelor plus aspirin for 1 month, followed by ticagrelor monotherapy for 23 months versus aspirin plus clopidogrel or ticagrelor for 12 months, followed by aspirin monotherapy for 12 months after implantation of a drug-eluting stent: a multicenter, open-label, randomized superiority trial

Original Research 2018 Oct 23.

JOURNAL:Catheter Cardiovasc Interv. Article Link

In vitro flow and optical coherence tomography comparison of two bailout techniques after failed provisional stenting for bifurcation percutaneous coronary interventions

Adjedj J, Picard F, Mogi S et al. Keywords: Reversed String; T and protrusion; bifurcation angle; in vitro bifurcation models; provisional T-stenting

ABSTRACT


OBJECTIVES - To evaluate, in vitro, SB stenting techniques after failed provisional stenting. We aimed to compare flows and stent strut apposition of T and protrusion (TAP) versus Reversed String (RS) techniques using a flow simulator, optical coherence tomography (OCT) using silicon bifurcation phantoms with different bifurcation angulations.

 

BACKGROUND -  While bifurcation coronary artery stenoses are preferably treated with provisional T-stenting strategy, the preferred bailout two stents technique to treat the side branch remains unclear.

 

METHODS AND RESULTS - Eleven 30°-angle and ten 60°-angle bifurcatio nphantoms were used. After performing provisional stenting, TAP and RS techniques were compared in six phantoms with 30° and five with 60° angles. Flow measurement was performed using absolute coronary flow and particle image velocimetry techniques. Strut apposition was evaluated using OCT. Flow analyses showed that disturbed flow regions were observed in the vicinity of floating struts protruded into the lumen both regardless of TAP and RS techniques. OCT analysis showed a higher proportion of floating struts protruding into the main branch with TAP compared to RS, respectively (13% vs. 1%; P <0.001) in both angles.

 

CONCLUSIONS - RS reduces the proportion of floating struts protruding into the main branch compared to TAP, at comparable flow rates. Clinical studies are needed to evaluate feasibility and potential clinical benefit of this technique.

 

© 2018 Wiley Periodicals, Inc.