中国心力衰竭诊断和治疗指南2018 Angiotensin–neprilysin inhibition versus enalapril in heart failure Diuretic Therapy for Patients With Heart Failure JACC State-of-the-Art Review Randomized Trial of Stents Versus Bypass Surgery for Left Main Coronary Artery Disease: 5-Year Outcomes of the PRECOMBAT Study Novel percutaneous interventional therapies in heart failure with preserved ejection fraction: an integrative review The spectrum of heart failure: value of left ventricular ejection fraction and its moving trajectories Frequency, predictors, and prognosis of ejection fraction improvement in heart failure: an echocardiogram-based registry study Effect of Intravascular Ultrasound-Guided Drug-Eluting Stent Implantation: Five-Year Follow-Up of the IVUS-XPL Randomized Trial Sildenafil for improving outcomes in patients with corrected valvular heart disease and persistent pulmonary hypertension: a multicenter, double-blind, randomized clinical trial Impact of the complexity of bifurcation lesions treated with drug-eluting stents: the DEFINITION study (Definitions and impact of complEx biFurcation lesIons on clinical outcomes after percutaNeous coronary IntervenTIOn using drug-eluting steNts)

Original Research 2018 Oct 23.

JOURNAL:Catheter Cardiovasc Interv. Article Link

In vitro flow and optical coherence tomography comparison of two bailout techniques after failed provisional stenting for bifurcation percutaneous coronary interventions

Adjedj J, Picard F, Mogi S et al. Keywords: Reversed String; T and protrusion; bifurcation angle; in vitro bifurcation models; provisional T-stenting

ABSTRACT


OBJECTIVES - To evaluate, in vitro, SB stenting techniques after failed provisional stenting. We aimed to compare flows and stent strut apposition of T and protrusion (TAP) versus Reversed String (RS) techniques using a flow simulator, optical coherence tomography (OCT) using silicon bifurcation phantoms with different bifurcation angulations.

 

BACKGROUND -  While bifurcation coronary artery stenoses are preferably treated with provisional T-stenting strategy, the preferred bailout two stents technique to treat the side branch remains unclear.

 

METHODS AND RESULTS - Eleven 30°-angle and ten 60°-angle bifurcatio nphantoms were used. After performing provisional stenting, TAP and RS techniques were compared in six phantoms with 30° and five with 60° angles. Flow measurement was performed using absolute coronary flow and particle image velocimetry techniques. Strut apposition was evaluated using OCT. Flow analyses showed that disturbed flow regions were observed in the vicinity of floating struts protruded into the lumen both regardless of TAP and RS techniques. OCT analysis showed a higher proportion of floating struts protruding into the main branch with TAP compared to RS, respectively (13% vs. 1%; P <0.001) in both angles.

 

CONCLUSIONS - RS reduces the proportion of floating struts protruding into the main branch compared to TAP, at comparable flow rates. Clinical studies are needed to evaluate feasibility and potential clinical benefit of this technique.

 