Predictors of Left Main Coronary Artery Disease in the ISCHEMIA Trial Optical coherence tomography and C-reactive protein in risk stratification of acute coronary syndromes High-Resolution Cardiac Magnetic Resonance Imaging Techniques for the Identification of Coronary Microvascular Dysfunction Long-term secondary prevention of cardiovascular disease with a Mediterranean diet and a low-fat diet (CORDIOPREV): a randomised controlled trial Intravascular optical coherence tomography Patient and Hospital Characteristics of Mitral Valve Surgery in the United States Diagnostic Performance of Angiogram-Derived Fractional Flow Reserve: A Pooled Analysis of 5 Prospective Cohort Studies Angiography Alone Versus Angiography Plus Optical Coherence Tomography to Guide Percutaneous Coronary Intervention Outcomes From the Pan-London PCI Cohort Active and Passive Vaccination for Pulmonary Arterial Hypertension: A Novel Therapeutic Paradigm New Volumetric Analysis Method for Stent Expansion and its Correlation With Final Fractional Flow Reserve and Clinical Outcome An ILUMIEN I Substudy

Original Research 2018 Oct 23.

JOURNAL:Catheter Cardiovasc Interv. Article Link

In vitro flow and optical coherence tomography comparison of two bailout techniques after failed provisional stenting for bifurcation percutaneous coronary interventions

Adjedj J, Picard F, Mogi S et al. Keywords: Reversed String; T and protrusion; bifurcation angle; in vitro bifurcation models; provisional T-stenting

ABSTRACT


OBJECTIVES - To evaluate, in vitro, SB stenting techniques after failed provisional stenting. We aimed to compare flows and stent strut apposition of T and protrusion (TAP) versus Reversed String (RS) techniques using a flow simulator, optical coherence tomography (OCT) using silicon bifurcation phantoms with different bifurcation angulations.

 

BACKGROUND -  While bifurcation coronary artery stenoses are preferably treated with provisional T-stenting strategy, the preferred bailout two stents technique to treat the side branch remains unclear.

 

METHODS AND RESULTS - Eleven 30°-angle and ten 60°-angle bifurcatio nphantoms were used. After performing provisional stenting, TAP and RS techniques were compared in six phantoms with 30° and five with 60° angles. Flow measurement was performed using absolute coronary flow and particle image velocimetry techniques. Strut apposition was evaluated using OCT. Flow analyses showed that disturbed flow regions were observed in the vicinity of floating struts protruded into the lumen both regardless of TAP and RS techniques. OCT analysis showed a higher proportion of floating struts protruding into the main branch with TAP compared to RS, respectively (13% vs. 1%; P <0.001) in both angles.

 

CONCLUSIONS - RS reduces the proportion of floating struts protruding into the main branch compared to TAP, at comparable flow rates. Clinical studies are needed to evaluate feasibility and potential clinical benefit of this technique.

 