Association of Sustained Blood Pressure Control with Multimorbidity Progression Among Older Adults 中国肺动脉高压诊断与治疗指南(2021版) 5-Year Outcomes Comparing Surgical Versus Transcatheter Aortic Valve Replacement in Patients With Chronic Kidney Disease 5-Year Outcomes After TAVR With Balloon-Expandable Versus Self-Expanding Valves: Results From the CHOICE Randomized Clinical Trial Stress Echocardiography and PH: What Do the Findings Mean? Randomized trial of simple versus complex drug-eluting stenting for bifurcation lesions: the British Bifurcation Coronary Study: old, new, and evolving strategies Short Length of Stay After Elective Transfemoral Transcatheter Aortic Valve Replacement Is Not Associated With Increased Early or Late Readmission Risk Extracellular Myocardial Volume in Patients With Aortic Stenosis Change in Kidney Function and 2-Year Mortality After Transcatheter Aortic Valve Replacement von Willebrand Factor and Management of Heart Valve Disease: JACC Review Topic of the Week

Clinical TrialVolume 6, Issue 10, October 2018

JOURNAL:JACC: Heart Failure Article Link

A Randomized Controlled Trial to Evaluate the Safety and Efficacy of Cardiac Contractility Modulation

Abraham WT, Kuck KH, FIX-HF-5 Investigators and Coordinators. Keywords: heart failure; exercise tolerance; peak Vo2; QRS duration; quality of life

ABSTRACT



OBJECTIVES - This study sought to confirm a subgroup analysis of the prior FIX-HF-5 (Evaluate Safety and Efficacy of the OPTIMIZER System in Subjects With Moderate-to-Severe Heart Failure) study showing that cardiac contractility modulation (CCM) improved exercise tolerance (ET) and quality of life in patients with ejection fractions between 25% and 45%.


BACKGROUND - CCM therapy for New York Heart Association (NYHA) functional class III and IV heart failure (HF) patients consists of nonexcitatory electrical signals delivered to the heart during the absolute refractory period.


METHODS - A total of 160 patients with NYHA functional class III or IV symptoms, QRS duration <130 ms, and ejection fraction 25% and 45% were randomized to continued medical therapy (control, n = 86) or CCM (treatment, n = 74, unblinded) for 24 weeks. Peak Vo2 (primary endpoint), Minnesota Living With Heart Failure questionnaire, NYHA functional class, and 6-min hall walk were measured at baseline and at 12 and 24 weeks. Bayesian repeated measures linear modeling was used for the primary endpoint analysis with 30% borrowing from the FIX-HF-5 subgroup. Safety was assessed by the percentage of patients free of device-related adverse events with a pre-specified lower bound of 70%.


RESULTS - The difference in peak Vo2 between groups was 0.84 (95% Bayesian credible interval - 0.123 to 1.552) ml O2/kg/min, satisfying the primary endpoint. Minnesota Living With Heart Failure questionnaire (p < 0.001), NYHA functional class (p < 0.001), and 6-min hall walk (p = 0.02) were all better in the treatment versus control group. There were 7 device-related events, yielding a lower bound of 80% of patients free of events, satisfying the primary safety endpoint. The composite of cardiovascular death and HF hospitalizations was reduced from 10.8% to 2.9% (p = 0.048).


CONCLUSIONS - CCM is safe, improves exercise tolerance and quality of life in the specified group of HF patients, and leads to fewer HF hospitalizations. (Evaluate Safety and Efficacy of the OPTIMIZER System in Subjects With Moderate-to-Severe Heart Failure; NCT01381172).