Comparison of intravascular ultrasound versus angiography-guided drug-eluting stent implantation: a meta-analysis of one randomised trial and ten observational studies involving 19,619 patients Clinical trial design and rationale of the Multicenter Study of MagLev Technology in Patients Undergoing Mechanical Circulatory Support Therapy With HeartMate 3 (MOMENTUM 3) investigational device exemption clinical study protocol Efficacy and Safety of Dapagliflozin in Heart Failure With Reduced Ejection Fraction According to Age: Insights From DAPA-HF The year in cardiovascular medicine 2020: heart failure and cardiomyopathies Atrial Fibrillation and the Prognostic Performance of Biomarkers in Heart Failure Optical Frequency Domain Imaging Versus Intravascular Ultrasound in Percutaneous Coronary Intervention (OPINION Trial) Results From the OPINION Imaging Study A three-vessel virtual histology intravascular ultrasound analysis of frequency and distribution of thin-cap fibroatheromas in patients with acute coronary syndrome or stable angina pectoris H2FPEF Score for Predicting Future Heart Failure in Stable Outpatients With Cardiovascular Risk Factors When and how to use SGLT2 inhibitors in patients with HFrEF or chronic kidney disease Clinical impact of intravascular ultrasound guidance in drug-eluting stent implantation for unprotected left main coronary disease: pooled analysis at the patient-level of 4 registries

Clinical TrialVolume 6, Issue 10, October 2018

JOURNAL:JACC: Heart Failure Article Link

A Randomized Controlled Trial to Evaluate the Safety and Efficacy of Cardiac Contractility Modulation

Abraham WT, Kuck KH, FIX-HF-5 Investigators and Coordinators. Keywords: heart failure; exercise tolerance; peak Vo2; QRS duration; quality of life

ABSTRACT



OBJECTIVES - This study sought to confirm a subgroup analysis of the prior FIX-HF-5 (Evaluate Safety and Efficacy of the OPTIMIZER System in Subjects With Moderate-to-Severe Heart Failure) study showing that cardiac contractility modulation (CCM) improved exercise tolerance (ET) and quality of life in patients with ejection fractions between 25% and 45%.


BACKGROUND - CCM therapy for New York Heart Association (NYHA) functional class III and IV heart failure (HF) patients consists of nonexcitatory electrical signals delivered to the heart during the absolute refractory period.


METHODS - A total of 160 patients with NYHA functional class III or IV symptoms, QRS duration <130 ms, and ejection fraction 25% and 45% were randomized to continued medical therapy (control, n = 86) or CCM (treatment, n = 74, unblinded) for 24 weeks. Peak Vo2 (primary endpoint), Minnesota Living With Heart Failure questionnaire, NYHA functional class, and 6-min hall walk were measured at baseline and at 12 and 24 weeks. Bayesian repeated measures linear modeling was used for the primary endpoint analysis with 30% borrowing from the FIX-HF-5 subgroup. Safety was assessed by the percentage of patients free of device-related adverse events with a pre-specified lower bound of 70%.


RESULTS - The difference in peak Vo2 between groups was 0.84 (95% Bayesian credible interval - 0.123 to 1.552) ml O2/kg/min, satisfying the primary endpoint. Minnesota Living With Heart Failure questionnaire (p < 0.001), NYHA functional class (p < 0.001), and 6-min hall walk (p = 0.02) were all better in the treatment versus control group. There were 7 device-related events, yielding a lower bound of 80% of patients free of events, satisfying the primary safety endpoint. The composite of cardiovascular death and HF hospitalizations was reduced from 10.8% to 2.9% (p = 0.048).


CONCLUSIONS - CCM is safe, improves exercise tolerance and quality of life in the specified group of HF patients, and leads to fewer HF hospitalizations. (Evaluate Safety and Efficacy of the OPTIMIZER System in Subjects With Moderate-to-Severe Heart Failure; NCT01381172).