The conductive function of biopolymer corrects myocardial scar conduction blockage and resynchronizes contraction to prevent heart failure Chimney technique in a TAVR-in-TAVR procedure with high risk of left main artery ostium occlusion Outcomes 2 Years After Transcatheter Aortic Valve Replacement in Patients at Low Surgical Risk Dapagliflozin for treating chronic heart failure with reduced ejection fraction 2020 ACC/AHA Guideline for the Management of Patients With Valvular Heart Disease: A Report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines Anticoagulation in Concomitant Chronic Kidney Disease and Atrial Fibrillation: JACC Review Topic of the Week Cardiac Structural Changes After Transcatheter Aortic Valve Replacement: Systematic Review and Meta-Analysis of Cardiovascular Magnetic Resonance Studies Rivaroxaban Plus Aspirin Versus Aspirin in Relation to Vascular Risk in the COMPASS Trial Switching of Oral Anticoagulation Therapy After PCI in Patients With Atrial Fibrillation: The RE-DUAL PCI Trial Subanalysis Impact of epicardial adipose tissue on cardiovascular haemodynamics, metabolic profile, and prognosis in heart failure

Clinical TrialVolume 11, Issue 23, December 2018

JOURNAL:JACC Cardiovasc Interv. Article Link

Comparison of 2 Different Drug-Coated Balloons in In-Stent Restenosis: The RESTORE ISR China Randomized Trial

YD Chen, L Gao, Q Qin et al. Keywords: drug-coated balloon; drug-eluting stent(s); in-segment late loss; in-stent restenosis

ABSTRACT


OBJECTIVES - The aim of the present study was to evaluate the angiographic efficacy, clinical safety, and effectiveness of the Restore paclitaxel-coated balloon in a randomized trial designed to enable the approval of the new device in China.

 

BACKGROUND - Drug-coated balloon (DCB) angioplasty offers an effective treatment for in-stent restenosis. Restore is a new DCB with a SAFEPAX shellac-ammonium salt excipient that can avoid drug washing off during catheter delivery to the target lesion site.

 

METHODS -  In the noninferiority RESTORE ISR China (Compare the Efficacy and Safety of RESTORE DEB and SeQuent Please in Chinese Patient With Coronary In-stent Restenosis) trial, eligible patients with first occurrence of drug-eluting stent ISR were randomized to the Restore DCB or SeQuent Please DCB in a 1:1 ratio stratified by diabetes. Angiographic and clinical follow-up was planned at 9 months and 1 year, respectively, in all patients. The study was powered for the primary endpoint of 9-month in-segment late loss.

 

RESULTS - Between May 2016 and July 2017, a total of 240 subjects at 12 sites were randomized to either the Restore group (n = 120) or the SeQuent Please group (n = 120). Nine-month in-segment late loss was 0.38 ± 0.50 mm with Restore versus 0.35 ± 0.47 mm with SeQuent Please; the 1-sided 97.5% upper confidence limit of the difference was 0.17 mm, achieving noninferiority of Restore compared with SeQuent Please (p for noninferiority = 0.02). Both DCBs had similar 1-year rates of target lesion failure (13.3% vs. 12.6%; p = 0.87).

 

CONCLUSIONS - In this head-to-head randomized trial, the Restore DCB was noninferior to the SeQuent Please DCB for the primary endpoint of 9-month in-segment late loss. (Compare the Efficacy and Safety of RESTORE DEB and SeQuent Please in Chinese Patient With Coronary In-stent Restenosis; NCT02944890)