CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

Coronary CT Angiography in Patients With Non-ST-Segment Elevation Acute Coronary Syndrome Optical coherence tomography compared with intravascular ultrasound and with angiography to guide coronary stent implantation (ILUMIEN III: OPTIMIZE PCI): a randomised controlled trial Impact of Abnormal Coronary Reactivity on Long-Term Clinical Outcomes in Women Prevalence and Prognosis of Unrecognized Myocardial Infarction Determined by Cardiac Magnetic Resonance in Older Adults Clinical Characteristics and Outcomes of STEMI Patients With Cardiogenic Shock and Cardiac Arrest Acute Myocardial Injury in Patients Hospitalized With COVID-19 Infection: A Review Post-Discharge Bleeding and Mortality Following Acute Coronary Syndromes With or Without PCI Comparison of the Preventive Efficacy of Rosuvastatin Versus Atorvastatin in Post-Contrast Acute Kidney Injury in Patients With ST-segment Elevation Myocardial Infarction Undergoing Percutaneous Coronary Intervention Geometry as a Confounder When Assessing Ventricular Systolic Function: Comparison Between Ejection Fraction and Strain Prognostically relevant periprocedural myocardial injury and infarction associated with percutaneous coronary interventions: a Consensus Document of the ESC Working Group on Cellular Biology of the Heart and European Association of Percutaneous Cardiovascular Interventions (EAPCI)

Clinical Trial2017 Oct 23;10(20):2029-2037

JOURNAL:JACC Cardiovasc Interv. Article Link

The SABRE Trial (Sirolimus Angioplasty Balloon for Coronary In-Stent Restenosis): Angiographic Results and 1-Year Clinical Outcomes

Verheye S, Vrolix M, Kumsars I et al. Keywords: drug-eluting balloon; extended release; in-stent restenosis; porous angioplasty balloon; sirolimus nanoparticle

ABSTRACT



OBJECTIVES - The aim of this first-in-human study was to assess the safety and effectiveness of the Virtue sirolimus-eluting balloon in a cohort of patients with in-stent restenosis (ISR).


BACKGROUND - Angioplasty balloons coated with the cytotoxic drug paclitaxel have been widely used for ISR treatment. The Virtue angioplasty balloon (Caliber Therapeutics, New Hope, Pennsylvania) delivers sirolimus in a nanoencapsulated liquid formulation. This clinical trial is the first to examine a sirolimus-eluting balloon for ISR.


METHODS - In this prospective, single-arm feasibility study at 9 European centers, 50 ISR patients were treated with the Virtue balloon. Angiographic measurements at 6 months are reported, along with 12-month clinicalfollow-up.


RESULTS - Procedural success in the intention-to-treat population was 100%. The primary safety endpoint was target lesion failure (TLF) (cardiac death, target vessel myocardial infarction, and clinically driven target lesion revascularization) assessed at 30 days (0%, n = 50). The primary performance endpoint was in-segment late lumen loss (LLL) at 6 months (0.31 ± 0.52 mm; n = 47). Secondary 6-month endpoints include binary restenosis (19.1%), diameter stenosis (30.3 ± 19.9%), and major adverse cardiac events (MACE) (10.2%, n = 49). In the 36-patient per-protocol population (excluding major protocol violations and previously stented ISR), LLL was 0.12 ± 0.33 mm at 6 months. Clinical outcomes at 1 year for the intention-to-treat group were 12.2% TLF and 14.3% MACE and for the per-protocol population were 2.8% TLF and 2.8% MACE.


CONCLUSIONS - This first-in-human study showed excellent procedural success for the Virtue sirolimus-eluting angioplasty balloon, 6-month LLL rates in line with current stent-free ISR treatment options, and clinicaloutcomes that warrant further evaluation in dedicated randomized studies.


Copyright © 2017 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.