CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

A Randomized Trial to Assess Regional Left Ventricular Function After Stent Implantation in Chronic Total Occlusion The REVASC Trial Guidelines in review: Comparison of the 2014 AHA/ACC guideline for the management of patients with non-ST-elevation acute coronary syndromes and the 2015 ESC guidelines for the management of acute coronary syndromes in patients presenting without persistent ST-segment elevation A Randomized Trial Comparing the NeoVas Sirolimus-Eluting Bioresorbable Scaffold and Metallic Everolimus-Eluting Stents Efficacy and safety of rosuvastatin vs. atorvastatin in lowering LDL cholesterol : A meta-analysis of trials with East Asian populations Qualitative Methodology in Cardiovascular Outcomes Research: A Contemporary Look Advances in Coronary No-Reflow Phenomenon-a Contemporary Review Long-Term Effect of Ultrathin-Strut Versus Thin-Strut Drug-Eluting Stents in Patients With Small Vessel Coronary Artery Disease Undergoing Percutaneous Coronary Intervention: A Subgroup Analysis of the BIOSCIENCE Randomized Trial Management of two major complications in the cardiac catheterisation laboratory: the no-reflow phenomenon and coronary perforations Clopidogrel or ticagrelor in acute coronary syndrome patients treated with newer-generation drug-eluting stents: CHANGE DAPT Homeostatic Chemokines and Prognosis in Patients With Acute Coronary Syndromes

Clinical Trial2018 Feb 3;391(10119):431-440.

JOURNAL:Lancet. Article Link

A sirolimus-eluting bioabsorbable polymer-coated stent (MiStent) versus an everolimus-eluting durable polymer stent (Xience) after percutaneous coronary intervention (DESSOLVE III): a randomised, single-blind, multicentre, non-inferiority, phase 3 trial

de Winter RJ, Katagiri Y, Serruys PW et al. Keywords: sirolimus-eluting bioabsorbable polymer-coated stent; everolimus-eluting durable polymer stent; non-inferiority

ABSTRACT


BACKGROUND - MiStent is a drug-eluting stent with a fully absorbable polymer coating containing and embedding a microcrystalline form of sirolimus into the vessel wall. It was developed to overcome the limitation of current durable polymer drug-eluting stents eluting amorphous sirolimus. The clinical effect of MiStent sirolimus-eluting stent compared with a durable polymer drug-eluting stents has not been investigated in a large randomised trial in an all-comer population.


METHODS - We did a randomised, single-blind, multicentre, phase 3 study (DESSOLVE III) at 20 hospitals in Germany, France, Netherlands, and Poland. Eligible participants were any patients aged at least 18 years who underwent percutaneous coronary intervention in a lesion and had a reference vessel diameter of 2·50-3·75 mm. We randomly assigned patients (1:1) to implantation of either a sirolimus-eluting bioresorbable polymer stent (MiStent) or an everolimus-eluting durable polymer stent (Xience). Randomisation was done by local investigators via web-based software with random blocks according to centre. The primary endpoint was a non-inferiority comparison of a device-oriented composite endpoint (DOCE)-cardiac death, target-vessel myocardial infarction, or clinically indicated target lesion revascularisation between the groups at 12 months after the procedure assessed by intention-to-treat. A margin of 4·0% was defined for non-inferiority of the MiStent group compared with the Xience group. All participants were included in the safety analyses. This trial is registered with ClinicalTrials.gov, number NCT02385279.


FINDINGS - Between March 20, and Dec 3, 2015, we randomly assigned 1398 patients with 2030 lesions; 703 patients with 1037 lesions were assigned to MiStent, of whom 697 received the index procedure, and 695 patients with 993 lesions were asssigned to Xience, of whom 690 received the index procedure. At 12 months, the primary endpoint had occurred in 40 patients (5·8%) in the sirolimus-eluting stent group and in 45 patients (6·5%) in the everolimus-eluting stent group (absolute difference -0·8% [95% CI -3·3 to 1·8], pnon-inferiority=0·0001). Procedural complications occurred in 12 patients (1·7%) in the sirolimus-eluting stent group and ten patients (1·4%) in the everolimus-eluting stent group; no clinical adverse events could be attributed to these dislodgements through a minimum of 12 months of follow-up. The rate of stent thrombosis, a safety indicator, did not differ between groups and was low in both treatment groups.


INTERPRETATION - The sirolimus-eluting bioabsorbable polymer stent was non-inferior to the everolimus-eluting durable polymer stent for a device-oriented composite clinical endpoint at 12 months in an all-comer population. MiStent seems a reasonable alternative to other stents in clinical practice.


FUNDING - The European Cardiovascular Research Institute, Micell Technologies (Durham, NC, USA), and Stentys (Paris, France).

 