CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

Comparison of 1-Year Pre- And Post-Transcatheter Aortic Valve Replacement Hospitalization Rates: A Population-Based Cohort Study Transcatheter Aortic Valve Replacement in Patients With Multivalvular Heart Disease The Science Underlying COVID-19: Implications for the Cardiovascular System Comparison of Early Surgical or Transcatheter Aortic Valve Replacement Versus Conservative Management in Low-Flow, Low-Gradient Aortic Stenosis Using Inverse Probability of Treatment Weighting: Results From the TOPAS Prospective Observational Cohort Study A Controlled Trial of Rivaroxaban After Transcatheter Aortic-Valve Replacement Impact of Staging Percutaneous Coronary Intervention in Left Main Artery Disease: Insights From the EXCEL Trial A Review of the Role of Breast Arterial Calcification for Cardiovascular Risk Stratification in Women Pulmonary arterial hypertension in congenital heart disease: an epidemiologic perspective from a Dutch registry Ten-Year All-Cause Death According to Completeness of Revascularization in Patients With Three-Vessel Disease or Left Main Coronary Artery Disease: Insights From the SYNTAX Extended Survival Study Anticoagulation After Surgical or Transcatheter Bioprosthetic Aortic Valve Replacement

Original Research2019 Feb 15;93(S1):772-778.

JOURNAL:Catheter Cardiovasc Interv. Article Link

Optical coherence tomography and intravascular ultrasound assessment of the anatomic size and wall thickness of a muscle bridge segment

Ye Z, Lai Y, Yao Y et al. Keywords: coronary perforation; myocardial bridging

ABSTRACT


OBJECTIVE - To use optical coherence tomography (OCT) and intravascular ultrasound (IVUS) in assessing myocardial bridging (MB) vessel size and wall thickness.

 

BACKGROUND - During stent implantation, MB is associated with complications, especially perforation.

 

METHODS - OCT and IVUS were performed in 56 patients with typical angiographic "milking" from November 2016 to May 2017. The vessel area and thickness in the MB segments and adjacent proximal and distal reference segments were measured and compared with eight normal left anterior descending (LAD) segment (no atherosclerosis in a segment that was at least 20 mm long and that began ~40 mm distal to the LAD ostium).

 

RESULTS - Compared with the reference vessel size distal to the MB segment (6.3 ± 1.8 mm2 ), the IVUS-measured size of the tunneled vessel during diastole was significantly smaller (6.0 ± 1.9 mm2 , p < 0.05) (remodeling index = 0.79 ± 0.18). The minimum intramyocardial arterial wall thickness was 0.16 ± 0.02 mm, significantly thinner than that of the mean reference (0.22 ± 0.03 mm, p < 0.001). The location of the thinnest arterial wall was in the distal and middle MB segments in 45 (80.4%) and 11 (19.6%) patients, respectively, and was not related to the degree of systolic compression or remodeling index. The walls of the middle and distal MB subsegments, but not of the proximal MB subsegment, were thinner than that of the comparison group of normal LADs.

 

CONCLUSION - The coronary vessel involved in an MB is both smaller and thinner than that of the adjacent non-MB segment. This may explain the increased frequency and severity of coronary perforation during stent implantation.

 

© 2019 Wiley Periodicals, Inc.