荐读文献

Age-Related Characteristics and Outcomes of Patients With Heart Failure With Preserved Ejection Fraction Atherosclerotic plaque with ultrasonic attenuation affects coronary reflow and infarct size in patients with acute coronary syndrome: an intravascular ultrasound study Sex- and Race-Related Differences in Characteristics and Outcomes of Hospitalizations for Heart Failure With Preserved Ejection Fraction Primary Prevention of Heart Failure in Women Mechanical circulatory support devices for acute right ventricular failure Timing of Intervention in Aortic Stenosis Myocardial bridging: contemporary understanding of pathophysiology with implications for diagnostic and therapeutic strategies Burden of Cardiovascular Diseases in China, 1990-2016: Findings From the 2016 Global Burden of Disease Study Guideline‐Directed Medical Therapy for Patients With Heart Failure With Midrange Ejection Fraction: A Patient‐Pooled Analysis From the KorHF and KorAHF Registries In-stent neoatherosclerosis: a final common pathway of late stent failure

Original Research2018 Dec 20;14(12):e1295-e1303.

JOURNAL:EuroIntervention. Article Link

Device specificity of vascular healing following implantation of bioresorbable vascular scaffolds and bioabsorbable polymer metallic drug-eluting stents in human coronary arteries: the ESTROFA OCT BVS vs. BP-DES study

de la Torre Hernandez JM, Gonzalo N, Otaegui I et al. Keywords: BVS vs. BP-DES; OCT; vascular healing

ABSTRACT

AIMS - We sought to compare vascular healing with bioresorbable everolimus-eluting vascular scaffolds (BVS) and drug-eluting stents with bioabsorbable polymers (BP-DES) at six and 12 months both implanted in the same patients.

 

METHODS AND RESULTS - This was a multicentre and prospective study including patients with at least two comparable lesions to treat. In every patient both BVS and BP-DES (SYNERGY, Orsiro or BioMatrix Flex) were implanted by lesion randomisation. Patients included were evaluated with optical coherence tomography at six or 12 months (2:1). Finally, 68 patients had an examination at six months and 27 patients at 12 months. The rates of uncovered struts at six months were 1.7±3.2% for BVS and 5.3±5.6% for BP-DES (p=0.0001), and at 12 months 0.48±0.72% and 4.8±5%, respectively (p=0.001). Rates of strut malapposition were significantly lower with BVS. There was no significant intra-patient correlation with BP-DES/BVS for endpoints. Evaginations were more frequent and larger with BVS. Discontinuities in BVS were observed in 19.4% at six months and 14.3% at 12 months.

 

CONCLUSIONS - Vascular healing with BVS and BP-DES could be more device-specific than patient-specific. At follow-up, BVS presented fewer uncovered or non-apposed struts than BP-DES but more frequent and larger evaginations. Discontinuities in BVS were relatively frequent at both time points.