CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

Rationale and design of a large-scale, app-based study to identify cardiac arrhythmias using a smartwatch: The Apple Heart Study The Evolution of β-Blockers in Coronary Artery Disease and Heart Failure (Part 1/5) Prior Balloon Valvuloplasty Versus Direct Transcatheter Aortic Valve Replacement: Results From the DIRECTAVI Trial Cardiac resynchronization therapy with a defibrillator (CRTd) in failing heart patients with type 2 diabetes mellitus and treated by glucagon-like peptide 1 receptor agonists (GLP-1 RA) therapy vs. conventional hypoglycemic drugs: arrhythmic burden, hospitalizations for heart failure, and CRTd responders rate Relationship Between Hospital Surgical Aortic Valve Replacement Volume and Transcatheter Aortic Valve Replacement Outcomes Negative Risk Markers for Cardiovascular Events in the Elderly Intravascular Ultrasound Assessment of In-Stent Restenosis in Saphenous Vein Grafts A Combined Optical Coherence Tomography and Intravascular Ultrasound Study on Plaque Rupture, Plaque Erosion, and Calcified Nodule in Patients With ST-Segment Elevation Myocardial Infarction: Incidence, Morphologic Characteristics, and Outcomes After Percutaneous Coronary Intervention Usefulness of intravascular ultrasound guidance in percutaneous coronary intervention with second-generation drug-eluting stents for chronic total occlusions (from the Multicenter Korean-Chronic Total Occlusion Registry) Meta-Analysis of Effectiveness and Safety of Transcatheter Aortic Valve Implantation Versus Surgical Aortic Valve Replacement in Low-to-Intermediate Surgical Risk Cohort

Original ResearchSeptember 2019

JOURNAL:JACC Cardiovasc Interv. Article Link

Long-Term Durability of Transcatheter Heart Valves: Insights From Bench Testing to 25 Years

J Sathananthan, M Hensey, U Landes et al. Keywords: durability; transcatheter aortic valve replacement; transcatheter heart valve

ABSTRACT


OBJECTIVES - This study assessed the long-term durability of nominally deployed transcatheter heart valves (THV) to 1 billion cycles (equivalent to 25 years) and non-nominal (overexpansion, underexpansion, and elliptical) THV deployments to 200 million cycles (equivalent to 5 years) with accelerated wear testing.


METHODS - SAPIEN 3 THVs, sized 20, 23, 26, and 29 mm were assessed. Nominally deployed THVs underwent hydrodynamic performance and mechanical durability as assessed with accelerated wear testing to 1 billion cycles. Magna Ease surgical valves were used as comparators. Durability of non-nominal THV deployments was tested to 200 million cycles. Valves were tested to International Standards Organization 5840:2013 standard.


RESULTS - THV durability was excellent for both the nominal and non-nominal THV deployments to 1 billion and 200 million cycles, respectively. At 1 billion cycles the regurgitant fraction for the 20-, 23-, 26-, and 29-mm SAPIEN 3 was 0.92 ± 0.47%, 1.29 ± 0.04%, 1.73 ± 0.46%, and 2.47 ± 0.15%, respectively. There was also excellent durability in the comparator Magna Ease valves. The regurgitant fraction of non-nominal overexpanded (20 mm, 4.36 ± 0.53; 23 mm, 7.68 ± 1.39; 26 mm, 6.80 ± 1.17; 29 mm, 9.00 ± 0.37), underexpanded (20 mm, 3.06 ± 0.28; 23 mm, 4.46 ± 0.45; 26 mm, 7.72 ± 0.48; 29 mm, 8.65 ± 2.01), and elliptical (20 mm, 3.30 ± 0.38; 23 mm, 6.13 ± 0.94; 26 mm, 6.77 ± 1.22; 29 mm, 8.72 ± 0.24) THVs were excellent at 200 million cycles.


CONCLUSIONS - Nominal SAPIEN 3 THVs demonstrated excellent durability, to an equivalent of 25-years wear. THV durability was similar to the comparator surgical valves tested. Non-nominal (overexpansion, underexpansion, and elliptical) THV deployments also had excellent durability to an equivalent of 5-years wear.