Transcatheter Aortic Valve Replacement in Low-Risk Patients With Symptomatic Severe Bicuspid Aortic Valve Stenosis Long-Term Outcomes After PCI or CABG for Left Main Coronary Artery Disease According to Lesion Location Transcatheter Versus Surgical Aortic Valve Replacement in Low-Risk Patients Right ventricular function and outcome in patients undergoing transcatheter aortic valve replacement 5-Year Outcomes After TAVR With Balloon-Expandable Versus Self-Expanding Valves: Results From the CHOICE Randomized Clinical Trial Expansion or contraction of stenting in coronary artery disease? Ticagrelor versus clopidogrel in patients with acute coronary syndromes Clinical Outcome After DK Crush Versus Culotte Stenting of Distal Left Main Bifurcation Lesions: The 3-Year Follow-Up Results of the DKCRUSH-III Study Ten-Year All-Cause Death According to Completeness of Revascularization in Patients With Three-Vessel Disease or Left Main Coronary Artery Disease: Insights From the SYNTAX Extended Survival Study Surgical ineligibility and mortality among patients with unprotected left main or multivessel coronary artery disease undergoing percutaneous coronary intervention

Clinical TrialSeptember 2019

JOURNAL:JACC Cardiovasc Interv. Article Link

Left Ventricular Rapid Pacing Via the Valve Delivery Guidewire in Transcatheter Aortic Valve Implantation

B Faurie, G Souteyrand, the EASY TAVI investigators. Keywords: left-ventricular stimulation; left-ventricular pacing; transcatheter aortic valve implantation; transcatheter aortic valve replacement

ABSTRACT


BACKGROUND - Rapid ventricular pacing is necessary to ensure cardiac standstill during transcatheter aortic valve implantation (TAVI).

 

OBJECTIVES - We investigated whether left ventricular (LV)-stimulation via a guidewire reduced procedure duration while maintaining efficacy and safety compared with standard right ventricular (RV)-stimulation.

 

 

METHODS - This is a prospective, multicenter, single-blinded, superiority, randomized controlled trial. Patients undergoing transfemoral TAVI with a Sapien valve (Edwards Lifesciences) were allocated to LV- or RV-stimulation. The primary endpoint was procedure duration. Secondary endpoints included efficacy, safety, and cost at 30 days. This trial is registered at clinicaltrials.gov (NCT02781896).

 

RESULTS - Between May 2017 and May 2018, 307 patients were randomised but 4 were excluded because they did not receive the intended treatment: 303 patients were analysed in the LV- (n=151) or RV-stimulation (n=152) groups. Mean procedure duration was significantly shorter in the LV-stimulation group (48.4±16.9 vs. 55.6±26.9 min, p=0.0013), with a difference of -0.12 (95% CI -0.20 to -0.05) in the log transformed procedure duration (p=0.0012). Effective stimulation was similar in the LV- and RV-stimulation groups: 124 (84.9%) vs. 128 (87.1%), p=0.60. Safety of stimulation was also similar in the LV- and RV-stimulation groups: procedural success occurred in 151 (100%) vs. 151 (99.3%) patients (p=0.99); 30-day MACE-TAVI occurred in 21 (13.9%) vs. 26 (17.1%) patients (p=0.44); fluoroscopy time was lower in the LV-stimulation group (13.48±5.98 vs. 14.60±5.59, p=0.02) as was cost (18,807±1,318 vs. 19,437±2,318, p=0.001).

 

CONCLUSIONS -  Compared with RV-stimulation, LV-stimulation during TAVI was associated with significantly reduced procedure duration, fluoroscopy time, and cost, with similar efficacy and safety.