Left Atrial Appendage Closure versus Non-Warfarin Oral Anticoagulation in Atrial Fibrillation: 4-Year Outcomes of PRAGUE-17 Frailty and Clinical Outcomes of Direct Oral Anticoagulants Versus Warfarin in Older Adults With Atrial Fibrillation: A Cohort Study Patent Foramen Ovale Attributable Cryptogenic Embolism With Thrombophilia Has Higher Risk for Recurrence and Responds to Closure Alcohol consumption, cardiac biomarkers, and risk of atrial fibrillation and adverse outcomes Role of endothelial dysfunction in determining angina after percutaneous coronary intervention: Learning from pathophysiology to optimize treatment Stretch-induced sarcoplasmic reticulum calcium leak is causatively associated with atrial fibrillation in pressure-overloaded hearts Systematic Review and Network Meta‐Analysis Comparing Bifurcation Techniques for Percutaneous Coronary Intervention Transcatheter Aortic Valve Implantation Represents an Anti-Inflammatory Therapy Via Reduction of Shear Stress-Induced, Piezo-1-Mediated Monocyte Activation Potential Candidates for Transcatheter Tricuspid Valve Intervention After Transcatheter Aortic Valve Replacement: Predictors and Prognosis Residual Shunt After Patent Foramen Ovale Closure and Long-Term Stroke Recurrence: A Prospective Cohort Study

Clinical TrialSeptember 2019

JOURNAL:JACC Cardiovasc Interv. Article Link

Left Ventricular Rapid Pacing Via the Valve Delivery Guidewire in Transcatheter Aortic Valve Implantation

B Faurie, G Souteyrand, the EASY TAVI investigators. Keywords: left-ventricular stimulation; left-ventricular pacing; transcatheter aortic valve implantation; transcatheter aortic valve replacement

ABSTRACT


BACKGROUND - Rapid ventricular pacing is necessary to ensure cardiac standstill during transcatheter aortic valve implantation (TAVI).

 

OBJECTIVES - We investigated whether left ventricular (LV)-stimulation via a guidewire reduced procedure duration while maintaining efficacy and safety compared with standard right ventricular (RV)-stimulation.

 

 

METHODS - This is a prospective, multicenter, single-blinded, superiority, randomized controlled trial. Patients undergoing transfemoral TAVI with a Sapien valve (Edwards Lifesciences) were allocated to LV- or RV-stimulation. The primary endpoint was procedure duration. Secondary endpoints included efficacy, safety, and cost at 30 days. This trial is registered at clinicaltrials.gov (NCT02781896).

 

RESULTS - Between May 2017 and May 2018, 307 patients were randomised but 4 were excluded because they did not receive the intended treatment: 303 patients were analysed in the LV- (n=151) or RV-stimulation (n=152) groups. Mean procedure duration was significantly shorter in the LV-stimulation group (48.4±16.9 vs. 55.6±26.9 min, p=0.0013), with a difference of -0.12 (95% CI -0.20 to -0.05) in the log transformed procedure duration (p=0.0012). Effective stimulation was similar in the LV- and RV-stimulation groups: 124 (84.9%) vs. 128 (87.1%), p=0.60. Safety of stimulation was also similar in the LV- and RV-stimulation groups: procedural success occurred in 151 (100%) vs. 151 (99.3%) patients (p=0.99); 30-day MACE-TAVI occurred in 21 (13.9%) vs. 26 (17.1%) patients (p=0.44); fluoroscopy time was lower in the LV-stimulation group (13.48±5.98 vs. 14.60±5.59, p=0.02) as was cost (18,807±1,318 vs. 19,437±2,318, p=0.001).

 

CONCLUSIONS -  Compared with RV-stimulation, LV-stimulation during TAVI was associated with significantly reduced procedure duration, fluoroscopy time, and cost, with similar efficacy and safety.