CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

Definitions and classifications of bifurcation lesions and treatment Propensity-Matched 1-Year Outcomes Following Transcatheter Aortic Valve Replacement in Low-Risk Bicuspid and Tricuspid Patients Defining cardiovascular toxicities of cancer therapies: an International Cardio-Oncology Society (IC-OS) consensus statement Beta-Blockers after Myocardial Infarction and Preserved Ejection Fraction Definition, classification and diagnosis of pulmonary hypertension Beta-Blockers after Myocardial Infarction and Preserved Ejection Fraction Viridans Streptococcal Biofilm Evades Immune Detection and Contributes to Inflammation and Rupture of Atherosclerotic Plaques Homocysteine metabolism as the target for predictive medical approach, disease prevention, prognosis, and treatments tailored to the person Endothelial ACKR3 drives atherosclerosis by promoting immune cell adhesion to vascular endothelium TRAP1 drives smooth muscle cell senescence and promotes atherosclerosis via HDAC3-primed histone H4 lysine 12 lactylation

Clinical TrialVolume 13, Issue 5, March 2020

JOURNAL:JACC Cardiovasc Interv. Article Link

Prior Balloon Valvuloplasty Versus Direct Transcatheter Aortic Valve Replacement: Results From the DIRECTAVI Trial

F Leclercq, P Robert, M Akodad et al. Keywords: balloon aortic valvuloplasty vs TAVR; device success; direct implantation

ABSTRACT


OBJECTIVES - The aim of this study was to evaluate device success of transcatheter aortic valve replacement (TAVR) using new-generation balloon-expandable prostheses with or without balloon aortic valvuloplasty (BAV).

 

BACKGROUND - Randomized studies are lacking comparing TAVR without BAV against the conventional technique of TAVR with BAV.

 

METHODS - DIRECTAVI (Direct Transcatheter Aortic Valve Implantation) was an open-label noninferiority study that randomized patients undergoing TAVR using the Edwards SAPIEN 3 valve with or without prior balloon valvuloplasty. The primary endpoint was the device success rate according to Valve Academic Research Consortium-2 criteria, which was evaluated using a 7% noninferiority margin. The secondary endpoint included procedural and 30-day adverse events.

 

RESULTS - Device success was recorded for 184 of 236 included patients (78.0%). The rate of device success in the direct implantation group (n = 97 [80.2%]) was noninferior to that in the BAV group (n = 87 [75.7%]) (mean difference 4.5%; 95% confidence interval: ?4.4% to 13.4%; p = 0.02 for noninferiority). No severe prosthesis-patient mismatch or severe aortic regurgitation occurred in any group. In the direct implantation group, 7 patients (5.8%) required BAV to cross the valve. Adverse events were related mainly to pacemaker implantation (20.9% in the BAV group vs. 19.0% in the direct implantation group; p = 0.70). No significant difference was found between the 2 strategies in duration of procedure, contrast volume, radiation exposure, or rate of post-dilatation.

 

CONCLUSIONS - Direct TAVR without prior BAV was noninferior to the conventional strategy using BAV with new-generation balloon-expandable valves, but without procedural simplification. BAV was needed to cross the valve in a few patients, suggesting a need for upstream selection on the basis of patient anatomy. (TAVI Without Balloon Predilatation [of the Aortic Valve] SAPIEN 3 [DIRECTAVI]; NCT02729519)