CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

Effect of Empagliflozin on Cardiovascular and Renal Outcomes in Patients With Heart Failure by Baseline Diabetes Status - Results from the EMPEROR-Reduced Trial Outcomes after drug-coated balloon treatment for patients with calcified coronary lesions Plaque progression assessed by a novel semi-automated quantitative plaque software on coronary computed tomography angiography between diabetes and non-diabetes patients: A propensity-score matching study Use of High-Risk Coronary Atherosclerotic Plaque Detection for Risk Stratification of Patients With Stable Chest Pain: A Secondary Analysis of the PROMISE Randomized Clinical Trial Contemporary use of drug-coated balloons in coronary artery disease: Where are we now? The performance of non-invasive tests to rule-in and rule-out significant coronary artery stenosis in patients with stable angina: a meta-analysis focused on post-test disease probability Impact of lesion complexity on peri-procedural adverse events and the benefit of potent intravenous platelet adenosine diphosphate receptor inhibition after percutaneous coronary intervention: core laboratory analysis from 10 854 patients from the CHAMPION PHOENIX trial A prospective natural-history study of coronary atherosclerosis Current Perspectives on Coronavirus Disease 2019 and Cardiovascular Disease: A White Paper by the JAHA Editors Multimodality imaging in cardiology: a statement on behalf of the Task Force on Multimodality Imaging of the European Association of Cardiovascular Imaging

Original ResearchVolume 13, Issue 6, March 2020

JOURNAL:JACC Cardiovasc Interv. Article Link

Coronary Access After TAVR

T Ochiai, T Chakravarty, SH Yoon et al. Keywords: aortic valve stenosis; coronary angiography; CAD; PCI; TAVR

ABSTRACT


OBJECTIVES - The aim of this study was to assess the incidence of unfavorable coronary access after transcatheter aortic valve replacement (TAVR) using post-implantation computed tomography (CT).

 

BACKGROUND - Real-world data regarding coronary access after TAVR assessed using post-implantation CT are scarce.

 

METHODS - Post-TAVR CT of 66 patients treated with Evolut R or Evolut PRO valves and 345 patients treated with SAPIEN 3 valves were analyzed. The distance from inflow of the transcatheter heart valve (THV) to the coronary ostia and the overlap between THV commissures and the coronary ostia were assessed. Coronary access was defined as unfavorable if the coronary ostium was below the skirt or in front of the THV commissural posts above the skirt in each coronary artery.

 

RESULTS - CT-identified features of unfavorable coronary access were observed in 34.8% (n = 23) for the left coronary artery and 25.8% (n = 17) for the right coronary artery in the Evolut R/Evolut PRO group, while those percentages were 15.7% (n = 54) for the left coronary artery and 8.1% (n = 28) for the right coronary artery in the SAPIEN 3 group. In the Evolut R/Evolut PRO group, 16 coronary engagements were performed after TAVR, while 64 coronary engagements were performed in the SAPIEN 3 group after TAVR. In an engagement-level analysis, the success rates of selective coronary engagement were significantly lower in patients with CT-identified features of unfavorable coronary access compared with those with favorable coronary access in both the Evolut R/Evolut PRO (0.0% vs. 77.8%; p = 0.003) and SAPIEN 3 (33.3% vs. 91.4%; p = 0.003) groups.

 

CONCLUSIONS - Coronary access may be challenging in a significant proportion of patients after TAVR. THVs with low skirt or commissure height and large open cells that are designed to achieve commissure-to-commissure alignment with the native aortic valve may facilitate future coronary access. (Assessment of Transcatheter and Surgical Aortic Bioprosthetic Valve Thrombosis and Its Treatment With Anticoagulation [RESOLVE]; NCT02318342)