Temporal Trends in Transcatheter Aortic Valve Replacement in France: FRANCE 2 to FRANCE TAVI Clinical applications of machine learning in the diagnosis, classification, and prediction of heart failure From Focal Lipid Storage to Systemic Inflammation Cardiovascular Considerations in Caring for Pregnant Patients: A Scientific Statement From the American Heart Association Primary Prevention of Sudden Cardiac Death Effect of Evolocumab on Complex Coronary Disease Requiring Revascularization Suture- or Plug-Based Large-Bore Arteriotomy Closure: A Pilot Randomized Controlled Trial Comparison of paclitaxel-eluting stents (Taxus) and everolimus-eluting stents (Xience) in left main coronary artery disease with 3 years follow-up (from the ESTROFA-LM registry) Universal Definition and Classification of Heart Failure: A Report of the Heart Failure Society of America, Heart Failure Association of the European Society of Cardiology, Japanese Heart Failure Society and Writing Committee of the Universal Definition of Heart Failure Permanent pacemaker use among patients with heart failure and preserved ejection fraction: Findings from the Acute Decompensated Heart Failure National Registry (ADHERE) National Registry

Original Research2015 Nov;8(13):1704-14.

JOURNAL:JACC Cardiovasc Interv. Article Link

Comparison of Stent Expansion Guided by Optical Coherence Tomography Versus Intravascular Ultrasound: The ILUMIEN II Study (Observational Study of Optical Coherence Tomography [OCT] in Patients Undergoing Fractional Flow Reserve [FFR] and Percutaneous Coronary Intervention)

Maehara A, Ben-Yehuda O, Stone GW et al. Keywords: intervention; intravascular ultrasound; optical coherence tomography; percutaneous coronary stent(s)

ABSTRACT


OBJECTIVES - The present study sought to determine whether optical coherence tomography (OCT) guidance results in a degree of stent expansion comparable to that with intravascular ultrasound (IVUS) guidance.


BACKGROUND - The most important predictor of adverse outcomes (thrombosis and restenosis) after stent implantation with IVUS guidance is the degree of stent expansion achieved.


METHODS - We compared the relative degree of stent expansion (defined as the minimal stent area divided by the mean of the proximal and distal reference lumen areas) after OCT-guided stenting in patients in the ILUMIEN(Observational Study of Optical Coherence Tomography [OCT] in Patients Undergoing Fractional Flow Reserve [FFR] and Percutaneous Coronary Intervention) (N = 354) and IVUS-guided stenting in patients in the ADAPT-DES (Assessment of Dual Antiplatelet Therapy With Drug-Eluting Stents) study (N = 586). Stent expansion was examined in all 940 patients in a covariate-adjusted analysis as well as in 286 propensity-matched pairs (total N = 572).


RESULTS - In the matched-pair analysis, the degree of stent expansion was not significantly different between OCT and IVUS guidance (median [first, third quartiles] = 72.8% [63.3, 81.3] vs. 70.6% [62.3, 78.8], respectively, p = 0.29). Similarly, after adjustment for baseline differences in the entire population, the degree of stent expansion was also not different between the 2 imaging modalities (p = 0.84). Although a higher prevalence of post-PCI stent malapposition, tissue protrusion, and edge dissections was detected by OCT, the rates of major malapposition, tissue protrusion, and dissections were similar after OCT- and IVUS-guided stenting.


CONCLUSIONS - In the present post-hoc analysis of 2 prospective studies, OCT and IVUS guidance resulted in a comparable degree of stent expansion. Randomized trials are warranted to compare the outcomes of OCT- and IVUS-guided coronary stent implantation.


Copyright © 2015 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.