Expansion or contraction of stenting in coronary artery disease? Randomized Evaluation of TriGuard 3 Cerebral Embolic Protection After Transcatheter Aortic Valve Replacement: REFLECT II Comparison of Outcomes of Percutaneous Coronary Intervention on Native Coronary Arteries Versus on Saphenous Venous Aorta Coronary Conduits in Patients With Low Left Ventricular Ejection Fraction and Impella Device Implantation Achieved or Attempted (from the PROTECT II Randomized Trial and the cVAD Registry) 2020 ACC Expert Consensus Decision Pathway on Management of Conduction Disturbances in Patients Undergoing Transcatheter Aortic Valve Replacement A Report of the American College of Cardiology Solution Set Oversight Committee MINOCA: a heterogenous group of conditions associated with myocardial damage Percutaneous Coronary Intervention Using Drug-Eluting Stents Versus Coronary Artery Bypass Grafting for Unprotected Left Main Coronary Artery Stenosis: A Meta-Analysis of Randomized Trials Evolution of antithrombotic therapy in patients undergoing percutaneous coronary intervention: a 40-year journey Impact of Staging Percutaneous Coronary Intervention in Left Main Artery Disease: Insights From the EXCEL Trial 5-Year Outcomes Comparing Surgical Versus Transcatheter Aortic Valve Replacement in Patients With Chronic Kidney Disease Transcatheter Aortic Valve Replacement: Role of Multimodality Imaging in Common and Complex Clinical Scenarios

Original Research2020 Apr 28.

JOURNAL:Eurointervention. Article Link

Double-Kissing Culotte Technique for Coronary Bifurcation Stenting - Technical evaluation and comparison with conventional double stenting techniques

GG Toth, V Sasi; D Franco et al. Keywords: technique improvement; culotte technique

ABSTRACT

AIMS - To assess, whether Culotte technique could be improved by an additional kissing dilation prior main branch (MB) stenting.

 

METHODS AND RESULTS - Double-kissing (DK) Culotte was compared to Culotte and DK-Crush techniques in bench model (n=24). Results were evaluated for stent apposition, for luminal opening and for flow dynamics.Total procedure duration of DK-Culotte was 18.3±3.4min, significantly lower than for DK-Crush (24.3±5.7min; p=0.015), however similar to Culotte (21.6±5.9min, p=0.104).In DK-Culotte overall rate of moderate (200-500mm) and significant (>500mm) malapposition was 2.1±1.9% and 0.4±0.2%; similar as compared to Culotte (3.7±3.8%, p=0.459 and 1.0±1.0%, p=0.517; respectively), and lower as compared to DK-Crush (8.1±2.5%, p<0.001 and 3.7±5.3%, p=0.002; respectively). Lower malapposition rate of DK-Culotte as compared to DK-Crush was due to less moderate and significant malapposition in proximal MB (0.0±0.0% vs. 14.0±7.6%, p<0.001 and 0.0±0.0% vs. 4.2±9.1%, p=0.026, respectively). Micro computed tomography did not show difference in luminal opening at proximal MB, distal MB or SB. There was no difference either in maximum shear rate or in areas of high shear or recirculation.

 

CONCLUSIONS - Bench tests data suggest that DK approach facilitates Culotte technique. Clinical validity and relevance remains to be confirmed in larger in vivo population.