CBS 2019
CBSMD教育中心
English

科学研究

科研文章

荐读文献

MITRA-FR vs. COAPT: Lessons from two trials with diametrically opposed results Single direct oral anticoagulant therapy in stable patients with atrial fibrillation beyond 1 year after coronary stent implantation Current Status and Future Prospects of Transcatheter Mitral Valve Replacement: JACC State-of-the-Art Review Gut microbiota dysbiosis promotes age-related atrial fibrillation by lipopolysaccharide and glucose-induced activation of NLRP3-inflammasome Detection of Device-Related Thrombosis Following Left Atrial Appendage Occlusion A Comparison Between Cardiac Computed Tomography and Transesophageal Echocardiography​: A Comparison Between Cardiac Computed Tomography and Transesophageal Echocardiography Left Atrial Appendage Occlusion during Cardiac Surgery to Prevent Stroke Role of local coronary blood flow patterns and shear stress on the development of microvascular and epicardial endothelial dysfunction and coronary plaque Management and outcomes of patients with left atrial appendage thrombus prior to percutaneous closure Transseptal puncture versus patent foramen ovale or atrial septal defect access for left atrial appendage closure Rivaroxaban Is Associated With Higher Rates of Gastrointestinal Bleeding Than Other Direct Oral Anticoagulants: A Nationwide Propensity Score–Weighted Study

Original Research2020 Aug;13(8):e009047.

JOURNAL:Circ Cardiovasc Interv . Article Link

Third-Generation Balloon and Self-Expandable Valves for Aortic Stenosis in Large and Extra-Large Aortic Annuli From the TAVR-LARGE Registry

G Armijo, GHL Tang, N Kooistra et al. Keywords: aortic valve stenosis; cohort studies; hemodynamics; humans; transcatheter aortic valve replacement.

ABSTRACT

BACKGROUND - Currently, 2 third-generation transcatheter valves, 29-mm Sapien-3 and 34-mm Evolut-R (ER), are indicated for large sized aortic annuli. We analyzed short and 1-year performance of these valves in patients with large (area 575 mm2or perimeter 85 mm) and extra-large (683 mm2or 94.2 mm) aortic annuli undergoing transcatheter aortic valve replacement.

 

METHODS - A total of 833 patients across 12 centers with symptomatic aortic stenosis and large aortic annuli underwent transcatheter aortic valve replacement with 29-mm Sapien-3 (n=640) or 34-mm ER (n=193). Clinical, anatomic, and procedural characteristics were collected, and Valve Academic Research Consortium-2 outcomes were reported.

 

RESULTS - Median aortic annulus area and perimeter were 617 mm2(591657) and 89.1 mm (87.092.1), respectively (704 mm2[689743] and 96.0 mm [94.597.9] in the subgroup of 124 patients with extra-large annuli). Overall device success was 94.3% (Sapien-3, 95.8% and ER, 89.3%;P=0.001), with a higher rate of significant paravalvular leak (P=0.004), second valve implantation (P=0.013), and valve embolization (P=0.009) in the ER group. Thirty-day and 1-year mortality was 2.4% and 9.2%, respectively, without differences between groups. Valve hemodynamics were excellent (mean gradient, 8.8±3.6 mm Hg; 3.3% rate of moderate-severe paravalvular leak) in the extra-large annulus, without differences compared with the large annulus group.

 

CONCLUSIONS - In patients with large and extra-large aortic annuli, transcatheter aortic valve replacement using 29-mm Sapien-3 and 34-mm ER is safe and feasible. Observed differences in clinical outcomes and hemodynamic performance may guide valve choice in this cohort of patients undergoing transcatheter aortic valve replacement.