2015 ESC Guidelines for the management of infective endocarditis: The Task Force for the Management of Infective Endocarditis of the European Society of Cardiology (ESC) Endorsed by: European Association for Cardio-Thoracic Surgery (EACTS), the European Association of Nuclear Medicine (EANM) Quality of Life after Everolimus-Eluting Stents or Bypass Surgery for Treatment of Left Main Disease Canagliflozin and Renal Outcomes in Type 2 Diabetes and Nephropathy Echocardiographic Screening for Pulmonary Hypertension in Congenital Heart Disease: JACC Review Topic of the Week Raising the Evidentiary Bar for Guideline Recommendations for TAVR: JACC Review Topic of the Week Intravascular Ultrasound Parameters Associated With Stent Thrombosis After Drug-Eluting Stent Deployment The impact of intravascular ultrasound guidance during drug eluting stent implantation on angiographic outcomes Successful bailout stenting strategy against lethal coronary dissection involving left main bifurcation Contribution of stent underexpansion to recurrence after sirolimus-eluting stent implantation for in-stent restenosis Histopathologic validation of the intravascular ultrasound diagnosis of calcified coronary artery nodules

Original Research2020 Aug;13(8):e009047.

JOURNAL:Circ Cardiovasc Interv . Article Link

Third-Generation Balloon and Self-Expandable Valves for Aortic Stenosis in Large and Extra-Large Aortic Annuli From the TAVR-LARGE Registry

G Armijo, GHL Tang, N Kooistra et al. Keywords: aortic valve stenosis; cohort studies; hemodynamics; humans; transcatheter aortic valve replacement.

ABSTRACT

BACKGROUND - Currently, 2 third-generation transcatheter valves, 29-mm Sapien-3 and 34-mm Evolut-R (ER), are indicated for large sized aortic annuli. We analyzed short and 1-year performance of these valves in patients with large (area 575 mm2or perimeter 85 mm) and extra-large (683 mm2or 94.2 mm) aortic annuli undergoing transcatheter aortic valve replacement.

 

METHODS - A total of 833 patients across 12 centers with symptomatic aortic stenosis and large aortic annuli underwent transcatheter aortic valve replacement with 29-mm Sapien-3 (n=640) or 34-mm ER (n=193). Clinical, anatomic, and procedural characteristics were collected, and Valve Academic Research Consortium-2 outcomes were reported.

 

RESULTS - Median aortic annulus area and perimeter were 617 mm2(591657) and 89.1 mm (87.092.1), respectively (704 mm2[689743] and 96.0 mm [94.597.9] in the subgroup of 124 patients with extra-large annuli). Overall device success was 94.3% (Sapien-3, 95.8% and ER, 89.3%;P=0.001), with a higher rate of significant paravalvular leak (P=0.004), second valve implantation (P=0.013), and valve embolization (P=0.009) in the ER group. Thirty-day and 1-year mortality was 2.4% and 9.2%, respectively, without differences between groups. Valve hemodynamics were excellent (mean gradient, 8.8±3.6 mm Hg; 3.3% rate of moderate-severe paravalvular leak) in the extra-large annulus, without differences compared with the large annulus group.

 

CONCLUSIONS - In patients with large and extra-large aortic annuli, transcatheter aortic valve replacement using 29-mm Sapien-3 and 34-mm ER is safe and feasible. Observed differences in clinical outcomes and hemodynamic performance may guide valve choice in this cohort of patients undergoing transcatheter aortic valve replacement.